ENVEAModel BERYL 92M FTUV -Ammonia (NH3) Analyzer (Fourier Transform technology)

FromENVEA
SHARE

Ultra-fast NH3 measurement (Ultra-fast response time : < 2 seconds). The use of the UV range for the measurement of NH3 has been specifically selected to avoid any interference with other major emission gases (CO, CO2, H2O etc..) that do not absorb in the UV.

Most popular related searches
  • Very high selectivity and maximum stability thanks to the UV Fourier Transform principle
  • Ultra-fast response time (< 2 sec)
  • High speed DSP processor
  • No risk of interferences or quenching effect due to  CO, CO2 or H2O
  • No internal converter, ultra-low maintenance
  • High lifetime of the lamp (> 1 year of continuous use)
  • Heated chamber (190°C)
  • Touch screen
  • Multilingual software
  • Automatic compensation of pressure and temperature variations
  • Engine exhaust gas monitoring
  • Emission monitoring (CEMS)
  • Process control (DeNOX SCR and SNCR)
  • Ranges: 0-100 / 0-500 ppm
  • Lower detectable limit: 0.4 ppm
  • Response time: < 2 sec (T90)
  • Zero drift: < 2% F.S. / 24 h
  • Span drift: 2 % F.S. / 24 h
  • Linearity: ± 1 % F.S.
  • Sample flow rate: 0.1 to 5 l/min
  • Heated block temperature: 190°C
  • Housing: Rack 19 ‘’ – 3U
  • Operating temperature: +5 to +45°C
  • Communication: RS 232
  • Protocol: AK and MODBUS