ENVEA

- Model MIR 9000 CLD - Multi-Gas IR-GFC analyzer

From

1

Standard Reference CLD method for low & ultra low NOx measurement, IR-GFC for CO, CO2, SO2, N2O, Nox, HF, HCl, TOC and O2 in a single analyzer. Fast and simultaneous measurements of up to 10 gases, available in 2 versions: NOx (CLD) and O2 (Paramagnetic) in 19’’ Rack or Tight box for low and ultra-low NOx monitoring. CO, CO2, SO2, HCl, HF, TOC, N2O (IR) + NOx (CLD) + O2 (Paramagnetic) in Tight box.

Multi-Gas IR-GFC analyzer (Infrared Gas Filter Correlation) with CLD option (Chemiluminescence Detector) designed to measure dry and corrosive samples

  • Over 4000 installations worldwide, covering many applications and industries
  • Unheated sample line thanks to the permeation drying technology used
  • Intrinsic security with on-board residual H2O measurements
  • On-board oxygen measure for environmental reporting
  • Interactive menu-driven software with real time graphic display for ease of operation
  • Highly accurate, excellent stability with automatic optical stability check
  • Automatic CO2 and cross interference correction
  • Built-in data logger for 7 additional parameters (flow, pressure, temperature or any other analogue input)
  • Compatible with various drying technologies such as a SEC sampling system or a high performance gas cooler

  • Cogeneration, Gas Turbines
  • Industrial Boilers
  • Furnaces
  • Power & Combustion

Lowest / Highest available ranges expressed in mg/m3 (may vary with your site conditions to be indicated on the Site Survey Form)

Technical Specifications

  • QAL1 to EN 15267-3 certified: YES
  • US EPA compliance: YES
  • Sampling technology: Cold/Dry Extraction
  • Sample transfert: non heated polytube
  • back flush function: YES
  • span injection: YES on analyser or at sample point
  • Built-In O2 option: YES
  • Serial link communication: RS232 / RS422
  • Other Communication: Ethernet
  • Analogue OUTPUTS: Optional
  • Analogue INPUTS: YES (x 4)
  • Dimensions (mm):
    • Tight box: 200 x 600 x 600 (D x W x H)
    • 19” rack: 490 x 483 x 177 (D x W x H)

Complete system would normally comprise of:

  • Sample extraction and conditioning probe (with integrated temperature, pressure and flow measurement)
  • Cold sample lines
  • Calibration module
  • Instrument air drying module
  • Acquisition and data management system (DAHS) may complement the system

Compliance with LCPD and WID Applications

  • QAL 1 certified by the MCERTs to EN 15267-3
  • QAL3 compliance as defined by EN 14181
  • U.S. EPA approved as compliant with 40 CFR 60 ET 75

Customer reviews

No reviews were found for ENVEA - Model MIR 9000 CLD - Multi-Gas IR-GFC analyzer. Be the first to review!