Erba Mannheim
  1. Companies
  2. Erba Mannheim
  3. Products
  4. Erba - Model ECL 760 - Seven Channel ...

ErbaModel ECL 760 -Seven Channel Fully Automated Random Access Analyzer

SHARE

ECL 760 is a seven channel fully automated random access analyzer for Complete Hemostasis. This system is ideal for mid-high range coagulation workload. It’s an advanced, fully automated random access analyzer for Complete Hemostasis. This seven channel coagulation analyzer redefines the benefits in its class. This system is ideal for mid-high range coagulation workload, especially for routine PT, aPTT and some rare parameters.

Most popular related searches
  • Interactive touch screen interface/ keyboard
  • Clotting Graph provides details of the clotting test
  • Cuvette and reagent management software
  • Twelve levels of QC with Levy-Jennings Plot- Identifies random and systematic errors and multirules.
  • Temperature stability at 37 degree C to perform the tests
  • Low Reagent Consumption
  • Windows software
  • CE mark

  • Clotting Tests such as PT, aPTT, Fibrinogen, Thrombin time, Factors, Lupus anticoagulant, Protein S can be performed using Light scatter principle
  • Protein C, Antithrombin III can be done by Chromogenic Principle
  • Tests like D Dimer can be performed using Immuno-turbidimetry principle

General Specifications:

  • 7 channel fully automated hemostasis instrument
  • It can perform 10 tests per profile
  • Sample id can be 14 digits numeric, self generating id or custom id
  • It is also possible to have continuous loading of samples 
  • It can run assays in single or duplicate (programmable per assay)
  • The System uses the much proven light scatter principle of detection

Tests principles:

  • Clotting (scattered light at 640nm)
  • Chromogenic (colorimetry at 405nm)
  • Immuno-turbidimetric (turbidimetry at 575nm)

Dimensions / Weight:

  • Dimension :660 x 580 x 510 mm
  • Weight : ~53 kg

Environmental requirements:

  • Operating Temperature : 17-32°C
  • Humidity: Max. 80% Relative Humidity, non-condensing

Power:

  • Power Supply : 100-240 VAC and 50/60 Hz
  • Power Consumption : 45 Watts
  • In-Rush Power : 150 VA or less

Calibrations:

  • Up to 6 calibration levels per assay
  • Storage of the active calibration curve

Quality Controls:

  • Up to 12 different control total can be assigned to any given assay.
  • Automatic Levey-Jennings plotting
  • Multi Rules
  • Storage of Quality Control results

Stored Results:

  • Independent storage allocation in the instrument memory for Quality Control (QC) and patient determinations
  • Up to100 results for QC determinations (run in duplicate)
  • Up to 100000 results for patients (run in duplicate)
  • New results automatically replace the oldest results of the appropriate category past the maximum storage allowance
  • Reaction curves are stored