TSI Incorporated
- Home
- Companies
- TSI Incorporated
- Products
- TSI - Model 3091 - Fast Mobility ...
TSI - Model 3091 -Fast Mobility Particle Size Spectrometer
FromTSI Incorporated
The FMPS spectrometer 3091 offers a total of 32 channels of resolution (16 channels per decade). It uses an electrical mobility measurement technique similar to that used in our SMPS™ spectrometer. However, instead of a CPC, the FMPS spectrometer uses multiple, low-noise electrometers for particle detection. This enables particle-size-distribution measurements in real time, providing the unique ability to visualize particle events and dynamic changes in particle size distribution.
The FMPS spectrometer operates at a high flow rate (10 L/min) to minimize diffusion losses of ultrafine and nanoparticles. It operates at ambient pressure to prevent evaporation of volatile particles, and it requires no consumables.
It is easy to transport, set up, and operate. Equipment be configured to measure single or multiple runs continuously for up to 12 hours in length. Its large, color VGA display and built-in control knob provide easy access to instrument features, set-up menus, and data displays.