SMTlabtech
  1. Companies
  2. SMTlabtech
  3. Products
  4. SMTlabtech - Model GC112 - GC112N Lab ...

SMTlabtechModel GC112 -GC112N Lab Analysis Instruments High Performance Gas Chromatograph With Cheap Price

SHARE
Products Description GC112N Lab Analysis Instruments High Performance Gas Chromatograph With Cheap Price GC112N gas chromatograph is suitable for trace detection of environmental protection, air and water pollution, analyzing poisons,monitoring and research, biochemistry, clinical application, pathology and virus research, food, petrochemicals, petroleum processing, oil analysis, geological , prospecting research, organic chemistry, synthesis, and quarantine, especially fit for power system, industrial and mining frontline laboratories and schools. if any question,please email to smtlabtech@163.com
Most popular related searches

if any question,please email to smtlabtech@163.com

or visit our website

Column temperature box
Content product
22L
Temperature control range
5℃ ~ 400℃ at room temperature
Temperature control accuracy
±0.1℃
Heating rate
0.1 ~ 60℃ / min
Program temperature rise order
9
Program heating repeatability
≤ 2%
Cooling way
open the door after
Cooling speed
≤10 mins (250℃ ~ 50℃)
Control software function
(GC112N only)
Column temperature box control
Detector control
Injector control
Map display
Sampler
Temperature control range
7℃ ~ 420℃ at room temperature
Temperature control method
independent temperature control
Carrier gas flow control mode
constant pressure
Number of simultaneous installations
3 at most
Type of injection unit
filling column, shunt
Split ratio
split ratio display
Cylinder pressure range
0 ~ 400kPa
Cylinder pressure control accuracy
0.1kPa
Flow setting range
H2 0 ~ 200ml / min
N2 0 ~ 150ml / min
Detector
FID, TCD optional
Temperature control
Max. 420℃
Number of simultaneous installations
2 at most
Ignition function
automatic
Hydrogen flame ionization detector (FID)
Detection limit
≤ 3×10 g/s (n-hexadecane)
Baseline noise
≤ 5× 10-14A
Baseline drift
≤ 6× 10-13A
Dynamic range
107
RSD
3% or less
Thermal conductivity detector (TCD)
Sensitivity
5000mV•mL/mg (n-cetane)
Baseline noise
≤ 0.05 mV
Baseline drift
≤ 0.15mV / 30min
Dynamic range
105
Supply voltage
AC220V±22V, 50Hz±0.5Hz
Power
3000W