GRIMM Aerosol Technik GmbH & Co. KG - a member of the Durag Group

GRIMMModel 1159-100 -Stand-alone Dilutor

SHARE

The 1159-100 is a diluter with a fix dilution ratio for a defined sample flow rate. It can be easily connected to any portable GRIMM aerosol spectrometer. The aerosol flow is split by a precision capillary into a sample and bypass flow. In the bypass flow all particles are removed by the built-in high-efficiency HEPA filter capsule. Downstream, both flows are mixed again. The actual dilution ratio is monitored by measuring the differential pressure over the capillary. No compressed air is needed.

Most popular related searches
  • no compressed air necessary for dilution
  • manual adjustment of best dilution fit when needed
  • small & portable
  • laboratory tests
  • exposition chambers
  • high dust loads
  • high particle concentrations
  • aerosol science
  • workplace monitoring
  • principle : capillary diluter
  • dilution media : internal HEPA filtered air
  • sample flow rate : 1.2 L/min
  • flow control : differential pressure sensor, manually set
  • dilution ratio : 1:100
  • capillary flow : 0.012 L/min
  • temperature range : 0 to +40 °C (32 to 104 °F)
  • pressure range : ±50 mbar
  • power supply : 110 – 220 VAC, 50 – 60 Hz