GRIMM Aerosol Technik Ainring GmbH & Co. KG - a member of the Durag Group

- Model 5420 SMPS+C - Scanning Mobility Particle Counter and Sizer

From

0

The GRIMM SMPS+C systems feature the Vienna-type DMA design (Winklmayr et al., 1991; Reischl et al., 1997), well known for highest size resolution and lowest particle diffusional losses – even for the smallest particles. The 19” rack-mounted versions of the SMPS+C systems, based on the GRIMM 5420 CPC, are operated at an aerosol inlet flow rate of 0.3 L/min. GRIMM offers a flexible design of the Vienna-type DMA with two electrodes of different length to accommodate a variety of experimental needs. To expand the size range, the GRIMM SMPS+C can be combined with a GRIMM Optical Particle Counter to build a Wide Range Aerosol Spectrometer (WRAS) that measures size distributions up to the particle size of 32µm.

  • particle size distribution from 5 – 1094 nm
  • two Vienna-type DMAs
  • aerosol flow rate = 0.3 L/min
  • sheath flow rate = 3 L/min
  • rugged, compact and reliable
  • fully automated use with our software
  • nalog inputs for additional sensors
  • anti-spill CPC-saturator design
  • operates with air and inert gases
  • comprehensive self-test for highest reliability

  • fundamental aerosol research
  • filter testing
  • environmental & climate studies
  • nanotechnology process monitoring
  • printer emission studies
  • inhalation & exposure studies
  • studies on atmospheric nucleation
  • studies on nanoparticle growth, coagulation & transport
  • engine exhaust studies
  • mobile aerosol studies
  • workplace monitoring

  • detector type : condensation particle counter (CPC)
  • working fluid : n-butanol (n-butyl alcohol)
  • max. concentration single count mode : 150 000 p/cm3
  • max. concentration photometric mode : 107 p/cm3
  • reproducibility :
    • > 95% for single count mode
    • > 90% for photometric mode
  • rise time t10 - t90 : < 3 s
  • size range :
    • M - DMA   5 – 350 nm
    • L -  DMA 10 - 1094 nm
  • size resolution : stepping mode: 45 - 255 channels scanning mode: 64 channels per decade logarithmic spacing

  • DMA dimensions :
    • Ri = 26 mm, Ro = 40 mm
    • L= 88 mm (M) or 350 mm (L)
  • output HV module : 5 – 10 000 V positive polarity negative polarity on request
  • internal pump : yes
  • flow rate sample air : 0.3 L/min
  • flow rate sheath air : 3.0 L/min
  • port for sensors : yes

  • ambient temperature : 10 to 40°C (50 to 104°F)
  • ambient humidity : 0 to 95% RH, non-condensing
  • absolute pressure range : 600 - 1100 mbar at full voltage range
  • power supply : 85 - 264 VAC, 47 - 440 Hz
  • power consumption : 80 - 130 W
  • interfaces : USB or RS-232
  • dimensions DMA (h x w x d) :
    • M – DMA: 23.4 x 14 x 15.6 cm (9.2 x 5.5 x 6.1 in)
    • L –  DMA: 47.8 x 14 x 15.6 cm (18.8 x 5.5 x 6.1 in)
  • weight DMA :
    • M – DMA: 5.7 kg (12.6 lbs)
    • L –  DMA: 7.9 kg (17.3 lbs)
  • dimensions CPC (h x w x d) : 40 x 25 x 29 cm (15.7 x 9.8 x 11.4 in)
  • weight CPC : 12.4 kg (27.3 lbs)

Customer reviews