H2S Medor - Hydrogen Sulfide (H2S) Analyzer
FromChromatotec
Analyser for continuous on-line monitoring of Hydrogen Sulphide.
Most popular related searches
H2S analyzer
H2S
sulfur compound analyzer
online gas chromatography
pneumatic valve
sulfur compounds analysis
hazardous area
permeation tube
remote monitoring
ATEX zone 1
H2S Medor is an online Gas Chromatograph dedicated to the analysis of H2S present in different matrices.
Two versions of the instrument exist :
- H2S Medor ppm (measuring at ppm levels)
- H2S Medor ppb (measuring at ppb levels)
- Internal pressure & temperature monitoring
- Sample loop injection is made by automatic valve
- H2S Medor uses a sulphur specific electrochemical cell for the detection of H2S
- Online, continuous sampling
- No interference from other compounds
- Low maintenance instrument – detector life may be up to 10 years
- Automatic Validation of results by Internal Permeation Tube
- Modifiable threshold alarms – High / Low alarm for deodorising processes
Chromatotec developed software system enables :
- Remote monitoring & injection control
- Full traceability with on board archiving of results and QC Set up and control of threshold alarms
- Export of data MODBUS / MGS1 / 4-20mA / 0-10V
- Multiple stream selector (up to 16 streams with one analyser)
- Modeling software with weather station for live odor/chemical impact
- Explosion proof version Exp or Exd for ATEX, IECEx zone 1 and 2 and also for CSA C1D2 group B , C & D T4
- Calculation modules (Average / Statistics / Odor unit OU e/m3 and more)
- Automatic data transfert through: Module for 4 outputs 4-20 mA (with 0 mA for instrument default) / 0 - 10 V / MODBUS RTU or TCPI IP or MGS1
- Nitrogen or air generator for safe and hazardous area
- 24 V DC power supply
- Electric selection valve to reduce air consumption
- Wall mounted box XXX934
- Liquid injection valve for LPG analysis
- Touch screen on the front panel for Exp version (only ATEX)
- External cylinder inlet for automatic calibration from external cylinder
- MFC for multipoint calibration from internal CALIB and/or from cylinder
- Zero method
Detection Limit:
- H2S MEDOR® ppm :
- H2S : 0.1 ppm (0.1417 mg/m3)
- H2S MEDOR® ppb :
- H2S : 1 ppb (1.39 μg/m3)
Range (adjustable depending on application):
- 0/10 or 0/100 or 0/1000 (ppb or ppm)
Relative Standard Deviation:
- RSD < 3 % on concentration over 48H
- RSD < 0.5% on retention time over 48H
Cycle Time:
- H2S 120 s for 0 - 3 ppm range
- H2S 300 s for higher range
Supervisor:
- Embedded computer Windows® based with LCD display
- 128 GB of Hardware storage on SSD memory
Communication:
- MODBUS RTU included in standard
- Ethernet, remote control
Gas supply:
- Carrier: Dry air or N2 (3 bar): < 4 ml/min
- Internal calibration: 50 ml/min for ppm range
- Sample inlet 1 bar and 160 ml/min sample flow
- Pneumatic valve 90 ml/commutation
Power supply:
- Main: 230 V / 115 V or 50/60 Hz
- Power: 24V (optional)
Electrical consumption:
- Average: 150 VA
Dimensions and weight:
- Rack: 19" (5U)
- Height: 222 mm
- Width: 482mm
- Depth: 600 mm
- Net weight: 22 Kgs