HORIBA Europe GmbH
  1. Companies
  2. HORIBA Europe GmbH
  3. Products
  4. HORIBA - Model EMIA-Pro - Carbon/Sulfur ...

HORIBAModel EMIA-Pro -Carbon/Sulfur Analyzer

SHARE

Enables everyone to become a professional analyst: The EMIA-Pro, Carbon/Sulfur Analyzer is based on HORIBA’s vast knowledge of Non-Dispersive Infrared (NDIR) technology and our wealth of experience acquired from our EMIA series. EMIA-Pro offers improved the cleaning efficiency, enhanced user-friendly software, durability, operability and maintenance to ensure efficient measurements and its shortened measurement cycle time accelerates repetitive measurements up to 200 times, thus improving development and manufacturing speed.

Most popular related searches

Long MTBM (Mean Time Between Maintenance)
The unique newly-developed cleaning mechanism (patent pending) clears the dust in the furnace, allowing the operator to continue to use the EMIA-Expert/Pro without maintenance cleaning for up to 200 measurements*.

* It depends on sample and measurement condition

High Throughput
EMIA Series can complete one measurement cycle, “measurement-display result-cleaning”, in 70 seconds (i.e. about 50 samples/hour)

Fast & Easy Maintenance
With the latest innovations the cleaning time for the EMIA Series instruments is typically half of that of our conventional models. The flat design around the crucible stand and simple design around the furnace also simplify user access for cleaning.

  • A critical focus was to ensure that the EMIA-Expert/Pro Series was “easy-to-use”. In particular, we focused on the layout, operating menus and functions.
  • We have included a self-diagnostic monitor checking capability to check the status of the connected device, an alarm function and we have expanded 3 navigators to facilitate daily operation.

Analysis Navigator 

  • This function uses the recommended best measurement flow and operating condition based on our experience. Explore more

Troubleshooting Navigator 

  • Automatic diagnosis system and troubleshooting navigator. Explore more

Maintenance Navigator

  • Periodic preventive maintenance work navigator by videos and photos. Explore more

Upgradable

The customer can extend the system’s functionality at any time, even after installation, depending upon their needs. Upgrades include: 

  • Carrier gas purifier for ultra-low concentration carbon measurement 
  • Auto-sampler (20 positions)
  • Halogen trap unit
  • Required sample amount: 1g±0.1g
  • Typical Meas. time: 70 seconds/cycle
    • Carbon
      • Meas. Range (m/m): 1.6 ppm - 6.0%
      • Blank accuracy: σn-1≦ 0.8 ppm
  • Sulfur
    • Meas. Range (m/m): 2.0 ppm - 1.0%
    • Blank accuracy: σn-1 ≦ 1.0 ppm
  • Sample meas. Accuracy
    • Carbon
      • σn-1 ≦ 2.0 ppm
      • or RSD≦1.0%
    • Sulfur
      • σn-1 ≦ 2.0 ppm
      • or RSD≦1.5%
  • Utility
    • Carrier gas: Oxygen (Purity; 99.5%, Pressure; 0.30-0.33MPa)
    • Operation gas: Nitrogen (Purity; 99.5%, Pressure; 0.35-0.38MPa)
    • Power: 200/220/240V, 50/60Hz, 5kVA
  • Dimensions
    • Main unit*: 500(W) x 725(D) x 710(H) mm
    • Mass Main unit: approximately 134 kg
  • Data processing and operation
    • USB data communication with PC Windows 8.1
    • Touch panel or Key board and mouse operation
  • Model name
    • EMIA-20P: Carbon/Sulfur
    • EMIA-21P: Carbon
    • EMIA-22P: Sulfur

Note) "ppm" is equal to mg/kg.
*Except piping parts, projection parts (D:79mm) and flow meter on the back side.