HORIBAModel HC-200F -Four-Wire Analyzer

SHARE

Suitable for fluoride ion monitoring in effluent from factories, e.g., those of semiconductor FPD and glass manufacturers. 0 to 10,000 mg/L wide measuring range (10-range switching).

Most popular related searches
  • Model : HC-200F
  • Combined sensor : Fluoride ion electrode 1009
  • Measurement range :
    • Concentration: 0 mg/L to 10000 mg/L (Display range: 0 mg/L to 11000 mg/L)
    • The selectable range: 10000 ,5000 ,2000 ,1000 ,500 ,200 ,100 ,50 ,20.0 ,10.0 mg/L
    • Temperature: 0°C to 100°C (Display range: -10°C to 110°C )
  • Transmission output :
    • Number of output: 2
    • (The negative terminals for transmission outputs are internally connected at the same electric potential)
    • Output type: 4 mA to 20 mA DC: input/output isolated type
    • Load resistance: Maximum: 900 Ω
  • Contact output :
    • Number of output: 3
    • Output type: No-voltage contact output
    • Contact type: Relay contact; SPDT (1c)
    • Output capacity: 250 VAC 3 A, 30 VDC 3 A (resistance load)
  • Cleaning output:
    • Number of output: 1
    • Output type: AC power control output ( Applied power supply voltage)
    • Contact type: Relay contact; SPST (1a)
    • Output contact capacity: 250 VAC 3 A, 30 VDC 3 A (resistance load)
    • External instrument: Solenoid valve for cleaning control
  • Contact input :
    • Number of inputs: 1
    • Contact type: No-voltage "a" contact for open collector
  • Communication function :
    • Communication type: RS-485
    • Signal type:
    • 2 wire system, isolated from the input circuit
    • Not isolated from transmission circuit
  • Temperature compensation : Applicable temperature element: Platinum resistor: 1kΩ (0°C), Positive relation resistor with temperature: 10kΩ(25°C), Temperature compensation range: 0 °C to 100°C
  • Calibration : Number of calibration points: Selectable from 1, and 2 points
  • Self-check: Calibration error, Electrode diagnostic error, Meter error.
  • Operating temperature range : -20°C to 55°C (without freeze)
  • Power supply : Power supply voltage range: 90 V to 264 VAC 50/60 Hz
  • Power consumption: 15VA(max)
  • Compatible standards : CE marking, FCC Rules
  • Structure :
    • Installation: Outdoor installation type
    • Installation method: Mounted on 50A pole or wall
    • Protection code: IP65 IEC60529, JIS C0920
    • Case material: Aluminum alloy (coated with epoxy-denatured melamine resin)
  • External dimensions : 180 (W) x 155 (H) x 115 (D) mm (excluding the fittings)
  • Mass : Main body: Approx. 3.5 kg; cover and fittings: Approx. 1 kg
  • Product name : Fluoride ion electrode
  • Model : 1009
  • Sample water conditions :
    • pH range:
    • 4 to 8 pH (at 0.2 mg/L)
    • 4 to 10 pH (at 20 mg/L)
    • 4 to 12 pH (at 2000 mg/L)
    • Temperature range: 0 to 40°C
    • Electric conductivity: Over 500 μS/cm
    • Flow rate range: 1 to 20 cm/sec.
  • Electrode structure :
    • Structure: Submergible type
    • Response film: Lanthanum fluoride
    • Comparative electrode: Ag/AgCl, Internal solution, KCL 3.3 mol/L, Replenishment type
    • Liquid junction: Ceramic double junction
    • Temperature compensation element: Pt1000Ω (0°C)
    • cable length: 5 m
  • External dimensions : 12Φ × 175 (L) mm(cable and projection excluded)
  • Mass : Approx. 200g