Knauer - Model 80P - High-Pressure Dosing Pump
From Pumps- Dosing Pumps
High-pressure dosing pump with 100 ml/min titanium pump head. Touchscreen control, automatic pump head recognition.
-
Most popular related searches
Products Details
The Preparative Pump 80P delivers stable and precise solvent flow for semi-preparative as well as preparative HPLC applications. By adding one or more additional pumps, binary, ternary or even quaternary high-pressure gradient systems can be easily configured. The pump can also be equipped with a binary or ternary gradient valve block to configure cost-effective low-pressure gradient systems. Exchangeable pump heads for flow rates up to 1 000 ml/min and pressures up to 40 MPa make it possible to easily optimize the pump to changing performance requirements and allow simple maintenance. Available in stainless steel or biocompatible titanium versions, the pump heads enable a wide range of applications.
Technical Data
- Max. flow rate 100 ml/min
- Flow rate increment 0.01 ml/min
- Maximum delivery pressure [psi] 5800 psi
- Maximum delivery pressure [bar] 400 bar
- Maximum delivery pressure [MPa] 40 MPa
- Pump head materials Titanium
- Maximum viscosity 1000 cP
- Liquid temperature range 4−60 °C (39.2−140 °F)
- Gradient Isocratic
- Leak management -
- Wetted materials Titanium, GFP (graphite fiber reinforced PTFE), aluminum oxide (Al2O3), PEEK, FKM, ruby, sapphire
- Note Pressure sensor membrane made of stainless steel; limited wetting in fluidic path
Detailed Information
- Best working conditions 1-80 ml/min
- Continuous working conditions 1-40 ml/min
- Flow rate accuracy ± 2 %
- Flow rate accuracy conditions measured at 5 - 50 % of flow range using ethanol/water 10:90
- Flow rate precision < 0.1 % RSD
- Flow rate precision conditions measured at 5 - 50 % of flow range using ethanol/water 10:90
- HPG: gradient accuracy ± 2 % (5 - 95 %, measured at 5 - 50 % of flow range, water/caffeine tracer)
- HPG: gradient precision 1 % RSD, based on retention time at constant room temperature
- LPG: gradient accuracy ± 3 % (5 - 95 %, measured at 5 - 50 % of the flow range, water/caffeine tracer)
- LPG: gradient precision 2 % RSD, based on retention time at constant room
- Active piston seal wash +
- Piston seal washing +
- Pump head inlet (standard) M8 x 1 (flat bottom)
- Pump head outlet (standard) M8 x 1 (coned)
Customer reviews