Labocon Systems Limited

LaboconModel LAAS-101 -Atomic Absorption Spectrophotometer

SHARE

Labocon Atomic Absorption Spectrophotometer LAAS-100 Series spectrophotometer, which offer superior base line stability of 0.002A/30min with advance optical system to ensure simple and reliable solution for trace elemental analysis.

Most popular related searches

  • Gas Control
    • Software controlled quick gas conversion and safetyprotection device used to analyse airacetylene flame and nitrous oxide-acetylene flame
  • Spray Chamber Metal Glass Spray
    • Burner Corrosion resistant titanium burners available for both air/acetylene and nitrous-oxide acetylene
  • Safety System
    • Safety interlock mechanism that monitors the burner type, burner fitted, gas pressures. Flame on/off sensors; Power failure automatic shutdown. Liquid trap system, pressure relief plug and flame shield to protect from UV radiation
  • Double beam
  • Automatic compensation of light source and wavelength drift caused by temperature variation and electronic circuit drift
  • Eliminates the need for pre-heating of cathode lamp to analyse the sample quickly
  • Gas path system equipped with precision pressure and current stabilizing devices to obtain stable flame at low noise
  • A very fine light beam passes through flame providing a precise analytical test with low characteristic concentration
  • Adopts total reflection system to eliminate color difference
  • Multi-functional analysis mode for flame absorption, flame emission, graphite furnace atomic absorption and hydride generation
  • Quick gas conversion and safety protection device used to analyse air-acetylene flame and nitrous oxide-acetylene flame

Labocon Atomic Absorption Spectrophotometer LAAS 100 series is widely used in analytical chemistry protocols, applications such as food and drug safety, clinical diagnostics, petrochemical, agricultural, environmental monitoring.

  • RS232 software
  • Graphite tubes
  • Graphite furnace system
  • Hydride generator
  • Nitrous oxide generator
  • Nitrous oxide regulator
  • Acetylene regulator
  • Recirculating cooling water system
  • Printer
  • Wavelength Range: 190-900 nm
  • Wavelength Accuracy: ≤±0.5 nm
  • Wavelength Repeatability: ≤0.3 nm (single direction)
  • Spectral Bandwidth: 0.2, 0.4, 0.7, 1.4, 2.4, 5.0 nm
  • Temperature Range: None
  • Working Humidity: None
  • Resolution: <40%
  • Baseline Stability: ≤0.004Abs/30min
  • Characteristic Concentration Limit: Cu: ≤0.04g/ml/1%
  • Detection Limit: Cu: 0.008g/ml
  • Background Calibration: Greater than 30 times
  • Overall Dimension: 1220x770x730 mm
  • Weight: 160 kg
  • Power: 660 W
  • Power Supply: 220V 50Hz
  • Catalog No.: 9523126164