Endress Hauser AG

- Model CA80CR - Chromate Analyzer System

From

0

Colorimetric system for the monitoring of industrial wastewater and optimization of process control. Liquiline System CA80CR offers highly accurate online measurement for optimized chromate removal in industrial wastewater. It supports you in achieving environmentally compliant electroplating and tanning processes. Like all Liquiline System analyzers, CA80CR enables plug & play of up to four Memosens sensors - minimizing the installation effort. Automatic calibration and cleaning plus advanced diagnostics with remote access ensure consistent process safety.

  • Standardized colorimetric measuring principle - diphenyle carbazide method - guarantees compliance with discharge regulations.

  • Reduced operating costs through automatic calibration and cleaning and low reagent consumption.

  • Simple maintenance: no tools required.

  • Advanced diagnostics with remote access for increased process safety.

  • Fast commissioning and plug & play thanks to Memosens technology and user-friendly operation.

  • Easy upgrade of functionality - even to a complete measuring station - simply by adding modules and connecting Memosens sensors. Reduces installation effort.

Liquiline System CA80CR supports you in improving the environmental performance of your processes by:

  • Monitoring and optimizing the cleaning capacity of industrial wastewater treatment plants.

  • Effluent monitoring in tanneries and electroplating plants.

  • Monitoring of membrane filtration.

  • Limit monitoring in drinking water applications.

  • Measuring principle

    Colorimetric

  • Characteristic

    Analyzer for Chromate (Chrome VI) in aqueous solutions

  • Size

    Cabinet

    793 x 530 x 417 mm

    31.22 x 20.87 x 16.42 in

    Stand housing

    1693 x 530 x 417 mm

    66.65 x 20.87 x 16.42 in

  • Design

    Open design, cabinet and stand housing

    High-Performance plastic ASA-PC, additional stand coated steel

  • Process temperature

    4 ... 40 °C (41 ... 104 °F)

  • Ambient temperature

    5 ... 40 °C (41 ... 104 °F)

  • Process pressure

    at atmospheric pressure, < 0.2 bar absolute

  • Sample flow rate

    min. 5ml/min (0,17 fl.oz/min)

  • Consistency of the sample

    Suspended solids content

    Turbitity < 50 NTU, aqueous, homogenized

  • Specials

    Easy upgrade to measuring station with up to four digital Memosens sensors

    Automatic calibration and cleaning

    User-configurable measuring, cleaning and calibration intervals

    Optional cooling module for longer reagent shelf life

    2 channel version optional

    User-definable measuring ranges

    Modular design for easily extensible functionality

    Digital communication for remote access

  • Application

    Limit value monitoring of Chromate in potable water and waste water.
    Common applications are industrial waste water Monitoring for tanneries and the galvanic industry.

  • Power supply

    100 ... 120 VAC / 200 ... 240 VAC ± 10%

    24 VDC ± 10%

    50 ± 1 or 60 ± 1,2 Hz

  • Output

    2x 0/4 ... 20 mA

    Modbus RS485/TCP (optional)

    Ethernet (optional)

    Alarmrelay

  • Input

    1 or 2 measuring channels

    1 ... 4 digital sensor inputs for sensors with Memosens protocol (optional)

  • Measurement range

    0,03 ... 2,5 mg/l Cr(VI)
    0,2 ... 5,0 mg/l Cr(VI)

  • Consumables

    Reagents and standard solutions CY80CR as well as cleaner CY800 are necessary for the operation. Regular maintenance is done with the parts of the maintenance kit CAV800.

Customer reviews