Endress Hauser AG

- Model CA80PH - Orthophosphate Analyzer System

From

0

Colorimetric system for wastewater monitoring and optimization of process control. Liquiline System CA80PH offers online orthophosphate measurement for precise dosing of precipitants in wastewater treatment. As all Liquiline System analyzers, it enables plug & play of up to 4 Memosens sensors - minimizing the installation effort. Automatic calibration and cleaning and the low consumption of reagents save you operating costs. Advanced diagnostics with remote access ensure process safety and support you in providing process documentation to the authorities.

  • Fast commissioning and plug & play thanks to Memosens technology and user-friendly operation

  • Easy upgrade of functionality - even to a complete measuring station - simply by adding modules and connecting Memosens sensors

  • Advanced diagnostics with remote access for increased process safety

  • Reduced operating costs through automatic calibration and cleaning and long reagent shelf life with cooling module

  • Available with both standardized measuring methods covering all applications: Molybdenum blue method and Molybdate Vanadate yellow method

Liquiline System CA80PH supports you in optimizing the cleaning capacity of wastewater treatment plants. It monitors and controls:

  • Precipitant dosing in aeration basins

  • Wastewater treatment plant outlets

In industrial applications, Liquiline System CA80PH is used for the monitoring of cooling water cycles.

  • Measuring principle

    Colorimetric

  • Characteristic

    Process analyzer for orthophosphate in aqueous solutions

    Comply with standard colorimetric measuring principle - molybdenum blue method - following DIN EN 1189 or molybdate vanadate method.

  • Size

    Cabinet
    793 x 530 x 417 mm
    31.22 x 20.87 x 16.42 in
    Stand housing
    1693

  • Design

    Open design, cabinet and stand housing
    High-Performance plastic ASA-PC, additional stand coated steel

  • Process temperature

    4 ... 40 °C (39.2 ... 104 °F)

  • Ambient temperature

    5 ... 40C ( 41 ... 104F)

  • Process pressure

    at atmospheric pressure, < 0.2 bar

  • Sample flow rate

    min. 5ml/min (0,17 fl.oz/min)

  • Consistency of the sample

    Suspended solids content

    Turbitity < 50 NTU, aqueous, homogenized

  • Application

    Monitoring and optimization of the cleaning capacity of municipal and industrial waste water treatment plants

  • Power supply

    100 ... 120 VAC / 200 ... 240 VAC ± 10%
    24 VDC ± 10%
    50 ± 1 oder 60 ± 1,2 Hz

  • Output

    2x 0/4 ... 20 mA
    Modbus RS485/TCP

  • Input

    1 or 2 measuring channel

    1 ... 4 digital sensor inputs for sensors with Memosens protocol (optional)

  • Measurement range

    0.05 ... 2.5 mg/l PO4-P (blue method)
    0.05 ... 10 mg/l PO4-P (blue method)
    0.5 ... 20 mg/l PO4-P (yellow method)
    0.5 ... 50 mg/l PO4-P (yellow method)

  • Consumables

    Reagents and standard solutions CY80H as well as cleaner CY800 ( blue method) are necessary for the operation. Regular maintenance is done with the parts of the maintenance kit CAV800.

Customer reviews