membraPure Aquinity² - Model P35 / P70 -Produces Pure Water and Ultra Pure Water System
- Analytical is a version with UV lamp for low TOC value. Life Science version comes with UV lamp and Ultrafiltration Module for low TOC and low endotoxin values.
- An optional integrated TOC sensor allows the user to monitor the TOC content permanentely online.
- Besides that the Aquinity² can be configured as bench top or bench-integrated system with external panel and wall mounted (with inclined display).
- The tank can be installed at the side or above the system or even underneath the bench.
- Upon request the tanks can be connected in series to reach a bigger volume of reservoir.
Produces Pure Water and Ultra Pure Water
- Compact housing
- Big touchscreen
- Remote service module
- Menus and wizards easy to use and mistake avoiding
- Fixed and flexible dispenser
- Easy access for cartridge exchange
Resistivity/Conductivity
- > 1 MOhm x cm | < 1 µS/cm; Type II
- 18.2 MOhm x cm | 0.055 µS/cm; Type I
Production rate
- max. 10 L/h, optionally 20 L/h
Total Organic Carbon (TOC)
- < 10 ppb (Reagent)
- < 5 ppb (Life Science)
- < 3 ppb (Analytical)
Perfect results with Water Purification Systems from membraPure
To deliver the best results for your analysis you need the highest water quality possible. With the “easy to operate” pure water and ultra pure water systems from membraPure you can select the model that cost efficiently produces laboratory-grade water. All our systems are Made in Germany.
Pure & Ultra Pure Water Purification Systems for all major lab applications
- Benchtop, wall mounted (with inclined display) and bench-integrated configurations
- Compact housing with small foot print
- Fast & easy cartridge replacement
- Easy-to-use software interface
- Intuitive software
- Up to 12 months trackable water quality values
- Integrated TOC monitoring
- Large touchscreen
- Flexible and lockable dispenser

Bench integrated configuration (BI version)
To save valuable benchtop space the system can be ordered as BI configuration. The system will be installed underneath the work bench and only the display and dispenser are mounted on the wall.
To save valuable benchtop space the system can be ordered as wall mounted configuration. the whole system can be mounted on the wall with inclined display.
Compact pump is required when feeding laboratory equipment such as autoclaves or laboratory dish washers.
The TOC monitoring during production and intermittent measurements during non-use periods allows to check the organic content in water continuously. The TOC value is measured between 1 and 999 ppb and can be shown in the display to monitor organic impurities.
Our dispenser allows the volumetric controlled dispensing of water with an increment of 0.1 L and a tap volume from 0.1 to 99 L. The system prevents overflow of containers and allows to dispense water without supervising.
Validation documents for water purification systems.
To be integrated into an ultra pure water system with a max, flow rate of 2 L/min. Point of use dispense arm with integrated control unit. Two joint to move it horizontally and vertically allowing to fill containers up to 55 cm height. Volumetric dosing of ultra pure water allows filling of containers without the attention of user, dosing eange 0.1 – 99 L, variable input and storage of values, after volume has been dispensed system goes back to standby.
- pure water + ultrapure water quality: < 2 µS/cm; Type II + 0.055 µS/cm; Type I
- resistivity: > 0.5 MegaOhm; Type II + 18.2 MegaOhm/cm; Type I
- total organic carbon (TOC): < 10 ppb Reagent, < 5 ppb Life Science, < 3 ppb Analytical
- dispensing flow rate: 2.0 L/min (1.5 L/min Life Science)
- production rate: 10 L/h, optionally 20 L/h
- bacteria: < 1 cfu/mL
- particulate: > 0.2 µm less than 1 particulate/mL
- pyrogen (endotoxins): < 0.001 EU/mL*
- RNAse: < 1 pg/mL*
- DNAse: < 5 pg/mL*
*Life Science
Feed Water Requirements
- type of feed water: tap water
- feed water conductivity: < 1400 µS/cm
- inlet pressure: 1.5 to 6 bar
- free chlorine: < 0.1 mg/L
- Silt Density Inded (SDI): < 3
- pH: 3 to 9
- temperature: 5 to 25 °C