Model DEN-1B -Densitometer (Suspension Turbidity Detector)

SHARE

Densitometer is designed for measurement of cell suspension`s turbidity in the range of 0.0–6.0 McFarland units (0 – 180×107 cells/ml).  Densitometers provide the opportunity to measure solution turbidity in a wider range (up to 15.0 McFarland units) however, it is necessary to remember that in this case the standard deviation values increase.

Most popular related searches

A densitometer is used for measurement of cell concentration (bacterial, yeast cells) during fermentation process, determination of microorganism sensitivity to antibiotics, microorganism identification using various test-systems, for measurement of absorption at the definite wavelength, as well as for quantitative estimation of concentration of colour solution, absorbing green light.

The operation principle is based on measurement of optical density with digital presentation of  results in McFarland units. The unit is calibrated at the factory (for operation with 16 mm diameter glass tubes) and keeps calibration without power supply. However, if necessary it is possible to calibrate the unit by 2–6 points in 0.0–6.0 McFarland unit range. Both commercial standards (e.g. produced by BioMerieux, Lachema, etc.) and the cell suspensions prepared in a laboratory can be used for calibration. Possibility to restore factory calibration settings*

Following calibration kits are available on request:

  • CKG16 for glass tubes with diameter 16 mm (latex particles);
  • CKG1802 for glass tubes with diameter 18 mm (BaSO4).

Standards CKG 1802 must be shaken before use. We recommend using vortex V-1 plus, which provides intensive mixing.

Two versions of the product are available:

  1. DEN-1 powered from external energy supply;
  2. DEN-1B powered both from external energy supply and from batteries (АА). Besides, DEN-1B operates with higher precision of measurements (up to 0,01 McF).
  • Light source LED
  • Measurement wavelength (λ) λ = 565 ±15 nm
  • Measurement range 0.00–15.00 McF
  • Display resolution 0.01 McF
  • Accuracy, of the full scale ±3%
  • Measurement time 1 s
  • Sample volume not less than 2 ml
  • Tube external diameter 16 mm (using A-16 adapter) or 18 mm (without adapter)
  • Possibility to restore factory calibration settings +
  • Display LCD
  • Overall dimensions (W×D×H) 165 × 115 × 75 mm
  • Weight 0.7 kg
  • Independent power supply 3 × AA batteries
  • Input current/power consumption 12 V, 7 mA / 0.1 W
  • External power supply Input AC 100–240 V, 50/60 Hz; Output DC 12 V
  • Standard set External power supply, A-16 and 3 × AA batteries