Hanna Instruments, Inc

Model DiST 5 - HI98311 -EC/TDS/ Temperature Tester

SHARE

When the original DiST (Dissolved Solids Tester) was first introduced, conductivity and total dissolved solids measurements became easy and affordable. The DiST’s ease of use, in combination with its affordability, made it the standard in EC and TDS measurement. Hanna continues the standard in EC and TDS testing with the DiST®5. This tester includes features such as a replaceable graphite electrode, adjustable TDS ratio, °C or °F measurement, Automatic Temperature Compensation (ATC) with adjustable β, battery level indicator, stability indicator, automatic shut-off and automatic calibration. The graphite conductivity electrode offers greater accuracy by resisting contamination by salt deposits in the sample. All of these features are packed in a floating, waterproof casing. This 3-in-1 tester is unmatched in EC, TDS, and temperature measurements.

Most popular related searches

When the original DiST (Dissolved Solids Tester) was first introduced, conductivity and total dissolved solids measurements became easy and affordable. The DiST’s ease of use, in combination with its affordability, made it the standard in EC and TDS measurement. Hanna continues the standard in EC and TDS testing with the DiST5.

Waterproof
Exposed temperature probe for fast response
Low range EC/TDS
0 to 3999 uS/cm 0 to 2000 ppm (mg/L)

  • Waterproof and designed to float
  • Automatic Temperature Compensation
  • HOLD button to freeze readings on the display
  • BEPS Alerts the user of low battery power that could adversely affect readings

Conductivity Range 0 to 3999 µS/cm
Conductivity Resolution1 µS/cm
Conductivity Accuracy±2% F.S.
TDS Range0 to 2000 ppm (mg/L)
TDS Resolution1 ppm (mg/L)
TDS Accuracy±2% F.S.
Temperature Range0.0 to 60.0°C / 32.0 to 140.0°F
Temperature Resolution0.1°C / 0.1°F
Temperature Accuracy±0.5°C /±1°F
EC/TDS Calibrationautomatic, one-point at: 1413 µS/cm, 1382 ppm (mg/L)
Temperature Compensationautomatic, with β adjustable from 0.0 to 2.4% / °C
TDS Conversion Factor0.45 to 1.00
Battery Type/Life1.5V (4) / approx. 100 hours of continuous use, auto-off after 8 minutes of non-use
Electrode/ProbeHI73311 replaceable EC/TDS graphite electrode (included)
Environment0 to 50°C (32 to 122°F); RH max 100%
Dimensions163 x 40 x 26 mm (6.4 x 1.6 x 1.0’’)
Weight100 g (3.5 oz.)
Ordering InformationHI98311 (DiST®5) is supplied with HI73311 EC/TDS probe, HI73128 probe removal tool, batteries and instructions.