TIGRET SP. Z O.O.
  1. Companies
  2. TIGRET SP. Z O.O.
  3. Products
  4. Model OdorPrep® V3 - Sample Collection

Model OdorPrep® V3 -Sample Collection

SHARE

The integrated system for continuous air quality monitoring and on-demand sampling of olfactory harassments. OdorPrep® V3 is equipped with an integrated control panel and comes with its own dedicated mobile applications for the remote control of the system, allowing the operation of all sample lines, or individually as required. The OdorPrep application is available via Play Store©3 sampling methods are available. Image for illustrative purposes. On-demand odor sampling method to EN13725 (Nalophan™ bags) The sampling line has an adjustable flow valve allowing sampling versatility. Duplicate or individual samples can be taken to allow for maximum flexibility and improved collection of the odorous air. The sampling lines are equipped with a protection system to preserve the sample following collection.

Most popular related searches

The sampling, storage and transport containers of the sample are rigid and sealed to avoid exposure to direct sunlight (two containers included for each sampling line), in order to minimize any chemical (photo)reactions and diffusion (as foreseen in the standard EN13725). The containers are returned for analysis in an accredited laboratory.

Sampling module on Nalophan bag™

 

Sampling method

Lung principle

Material

Nalophan™ bag (not included)

Nominal flow

8.5 l/min.

Collection tube

PVC stiff and opaque

Operating flow

1.2 – 5.6 l/min.   

Tube dimension (cm)   

82 H x 15 ID x 19 OD

Volumetric capacity

10 l

Weight

Approx. 5 Kg

 

On-demand sampling module on thermal or chemical desorption cartridge

The sampling line on cartridge (thermal desorption or sorbent tube) includes a high-quality vacuum pump and mass flow controller for ensuring the correct flow rate is maintained and a record of the total sampling volume is provided.

Sampling module on cartridge  

Sampling

Solid adsorption support (not included)

Operating flow

20 ml/min. – 1300 ml/min.

On-demand sampling module on Canister

The sampling line is equipped with a control unit for the remote opening and closing of the specialist valve, a dedicated digital pressure sensor monitors any vacuum losses while the sampler is operational.

The duration of the sampling depends on the restrictor used. Below are some examples of sampling durations depending on the restrictor and the Canister volume.

Critical orifice diameter of the restrictor vs. flow rate

Orifice Diameter

Flow Rate Range  

Canister Volume / Sampling Time  

(in.)

(mL/min.)

1L

3L

6L

15L

0.0008

0.5–2

24 hr.

48 hr.  

125 hr.  

0.0012

2–4

4 hr.

12 hr.

24 hr.

60 hr.  

0.0016

4–8

2 hr.

6 hr.

12 hr.

30 hr.

0.0020

8–15

1 hr.

4 hr.

8 hr.

20 hr.

0.0030

15–30

2 hr.

3 hr.

8 hr.

0.0060

30–80

1.5 hr.

4 hr.

0.0090

80–340

0.5 hr.

1 hr.

The canister sampling lines are independent. Sampling can be performed on several lines simultaneously or in sequence. The sampling unit is equipped with a temperature control system to allow the correct storage of the samples and the sampling temperature above the dew point, avoiding condensation. At the end of the sampling period a notification is sent to the operators so that collection of the canisters can be arranged.

Different configurations available for on-demand sampling are shown below.

Available sampling lines

Abbreviation

Rigid and opaque vacuum tube for sampling on Nalophan© bag

T

Sampling module on a thermal-chemical desorption cartridge

F

Sampling module on Canister

C

   
 

Configuration

I° Line

II° Line

III° Line

IV° Line

 
 

TT

Nalophan™

Nalophan™

X

X

 
 

FF

Cartridge

Cartridge

X

X

 
 

CC

Canister

Canister

X

X

 
 

TF

Nalophan™

Cartridge

X

X

 
 

TC

Nalophan™

Canister

X

X

 
 

FC

Cartridge

Canister

X

X

 
 

TFC

Nalophan™

Cartridge

Canister

X

 
 

TTF

Nalophan™

Nalophan™

Cartridge

X

 
 

FFC

Cartridge

Cartridge

Canister

X

 
 

CCF

Canister

Canister

Cartridge

X

 
 

CCFF

Canister

Canister

Cartridge

Cartridge

 
             

OdorPrep V3 – On-demand sampling system

Power supply: 220 VAC ; 24 VDC ; 50-60 Hz

Dimension (L x H x D): 82 cm  x 125 cm x  50 cm (sampling point excluded)

Communication: 3G / 4G (SIM card not included)

Protection: IP65 – Key box panel lock

Weight: 100 Kg

Transport: Cart with 4 swivel and self-locking wheels