SIGAS Measurement Engineering Corp.

NDIR multi-Gas Module Silarex

SHARE

The Silarex is one high-end type of multi-gas measurement sensor which based on the infrared absorption technology and comply with the highest standards in accuracy, reliability and efficiency. Due the consequent use of this optical principle the systems work without abrasion or chemical reactions and are nearly maintenance free. As basis the IR principle uses the individual absorption spectrum of the gas to be detected and specifies the concentration by the precise quantitative analysis. Because all measured gases absorb IR radiation in a different spectrum a so called “fingerprint” is given which ensures a selective characterization with less cross sensitivities. An integrated evaluation electronic delivers finally an easy to use concentration signal and compensates all temperature- and drift effects.



Most popular related searches

The Silarex housing is made with gold plating housing which can effectively reduce the noise of light source. And the gas Inlet and Outlet are fitted with gas line connectors to allow perfusion of the Silarex.  Tubing with an internal diameter of 3 mm and external diameter of 5 mm is needed to connect up to the measurement cell. Ensure that tubing is securely attached to the inlet and outlet connectors.

Meanwhile we have equipped Silarex with heated aluminium Housing to make the stable temperature of sensor, to reduce the fluctuation of measurement date due to the unstable environment temperature. Also with pressure sensor to reduce the gas flow influence. Silarex is the right product for all kind of industry stable gas measurement application.



Detection Principle NDIR (Dual Beam)
Measuring Range See List
Gas Supply Flow
Dimensions See Mechanical Data
Hose connection 3mm internal, 5mm outside
Heated Tem. Internal 42°C
Response time (T90) < 8sec @0.7 l/min (default)
Repeatability ≤ ±1 % FS
Linearity Error ≤ ±1 % FS ( or ≤ ±1.5 % FS depend on version)
Stability < ±2% [FS] @1000h
Operating Temp 0 °C up to 40 °C
Storage Temp -20 °C up to 60 °C
Ambient Humidity 0 % up to 95 % rel. Humidity (Non-condensing)
Ambient Air Pressure 800 ... 1150 mbar
Warm up Time < 2 minutes (Cold start)          < 30 minutes according to heat isolation and environment Tem.
Flow 0,2up to 1,0 L / min (const flow)
Gas Inlet Tem. 2 ... +42°C
Gas Inlet Humidity 5°C max ±1K (need cooler)
Diginal Output Signal       Modbus ASCII RTU TTL
2400-115.200 auto Baud/ auto Frame
Operation power 24 VDC  ±0.3V
Driven Current 240mA / max 1.5A during Heating
Power Consumption < 800 mW / max 36W during Heating