- Home
- Companies
- ECO MEDICS AG
- Products
- ECO MEDICS - Model CLD 88 - Nitric ...
ECO MEDICS - Model CLD 88 -Nitric Oxide Analyzer for Liquid NO Samples
The ANALYZER CLD 88 is designed for the precise measurement of nitric oxide in liquid samples, providing unparalleled sensitivity that enables the detection of NO concentrations down to parts per trillion. Its rapid response time and user-friendly design make it suitable for both standard applications and scientific research. When paired with a liquid purge vessel, this analyzer can accurately measure nitrites, nitrates, and nitrosothiols in biological fluids such as plasma, urine, and sera at pMol levels. The device's functionality is enhanced by the PowerChrom™️ software from EDAQ, which facilitates seamless data recording and analysis. This versatile setup can be adapted to measure NO from various sources, catering to biomedical, pharmaceutical, and laboratory applications. However, note that NO-calibration gases and the PowerChrom system with software and PC are sold separately. ECO MEDICS and EDAQ emphasize that specifications can change without prior notice.
The ANALYZER CLD 88 offers precise measurement of nitric oxide. The extremely high sensitivity and fast response time of the NO-analyzer allows the detection of concentrations in the range of parts per trillion. Ease of use guarantee flexibility for standard applications as well as scientific research. In combination with the liquid purge vessel the CLD 88 measures nitrites, nitrates and nitrosothiols in biological fluids like plasma, urine and sera down to pMol concentrations. The easy and accurate data recording and analysis is guaranteed by using the PowerChrom™️ software from EDAQ.
Fast and accurate
The ANALYZER CLD 88 is an extremely fast and accurate instrument with a detection range down to 0.06 ppb and a linearity of ±1% over the full scale.
User friendly software package
PowerChrom is a simple to use software for the collection, display and analysis of NO measurements.
Wide application spectrum
The setup can be easily adapted so that NO can be measured from many different sources. Applications are manifold and range from biomedical and pharmaceutical to laboratory applications.
NO measurement
- Measuring range – Gas samples: 0.1 – 5000 ppb
- Measuring range – Liquid samples: pico- to millimolar
- Min. detectable concentration: 0.06 ppb*
- Noise at zero-point (1σ): 0.03 ppb*
- Zero drift: <0.5 ppb per 6 h
- Linearity: ± 1% full scale
- Lag time: < 0.5 s
- Rise time (T90): < 100 ms
- Sample flow: selectable type 1 or 3 (110 or 330 ml/min)
- Sampling rate: 40 Hz internal, 10 Hz output
Interface : Analog output: 20 mA into 600 Ohm, galvanic isolated
General
- Temperature range: 10 – 40 °C
- Humidity tolerance: 5 – 95% rel. humidity (non-condensing)
- Ozone generator: internal, maintenance free (no external supply gas required)
- Ozone scrubber: internal, thermal, maintenance free
- Power consumption: 400 VA
- Supply voltage: 100 – 240 V / 50 – 60 Hz
- Weight: 24 kg
- Dimensions: H x W x D: 135 x 500 x 540 mm (5¼ x 19 x 17 “)
- System includes: ANALYZER CLD 88, sample tube type 3, mains cable and operator’s manual