Dwyer Instruments, LLC d/b/a DwyerOmega
  1. Companies
  2. Dwyer Instruments, LLC d/b/a DwyerOmega
  3. Products
  4. Omega - Model FMA5502A - Gas Mass flow ...

OmegaModel FMA5502A -Gas Mass flow Controllers with or Without Display

SHARE

The gas mass flow controllers of FMA5400A/5500A Series provide for controlling the flow of a wide variety of gases from 0-10 SCCM up to 100 SLM. Utilizing heat transfer through a heated tube to measure molecular gas flowrate, the FMA5400A/5500A provides measurement of direct gas mass flowrate, without the need to compensate for variations in gas temperature or pressure (within stated limits). They are available in an economical aluminium/brass construction for typical gas flows and a 316 SS construction for applications requiring more corrosion resistance. The FMA5400A series without integral display is supplied with both an analog 0-5 Vdc and 4 to 20 mA output for remote monitoring; the FMA5500A series has both an integral 3-1/2 digit display and analog output. The display is tiltable over 90 degrees for viewing convenience. The display is calibrated standard to read out directly in SCCM or SLM for nitrogen (other gas calibrations available on special order).

Most popular related searches
  • Flowmeter to read and Control Gas Mass Flow Without Temperature or Pressure Compensation
  • NIST Traceable Calibration
  • Available in Economical Aluminum or Corrosion-Resistant 316 SS
  • Tiltable LCD Display for Easy Reading

The FMA5400A/5500A mass flow controllers feature a built-in normally-closed electromagnetic valve for maintaining a constant flowrate regardless of variations in inlet or outlet pressures. The setpoint is controlled either locally via a potentiometer accessible through an access window in the case or remotely via an analog 0-5 Vdc or 4-20 mA signal (field selectable). The FMA5400A/5500A series require 12 to 26 Vdc power @ 800 mA maximum (12 to 15 Vdc sizes 43, 44 and 45 only), which can be supplied by the FMA545PW wall plug-in socket power supply. Model number FMA545C (supplied separately) provides a mating 15-pin “D” connector with 8 feet of shielded cable for accessing the analog output signals and power input connections. The LCD supplied with the FMA5500A series is connected to the lower electronics via a modular plug. The LCD can be remotely located by purchasing an FMA18RC remote cable assembly–you must then build your own assembly for panel mounting the LCD.

  • Accuracy: ±1.5% FS, including linearity over 15 to 25°C (59 to 77°F) and
  • 0.7 to 4.2 kg/cm² (10 to 60 psia) ±3% FS for units ≥ 100 SLM from o to 20% of range.
  • Repeatability: ±0.5% of full scale
  • Temp Coefficient: 0.15% FS/°C
  • Pressure Coefficient: 0.01% FS per psi (0.07 bar)
  • Maximum Pressure Drop: 50 psid
  • Response Time: 5 seconds to within ±2% of set flow rate over 25 to 100% FS
  • Max Gas Pressure: 70 kg/cm² gauge (1000 psig)(For units up to 100 SLM); 35 kg/cm² gauge (500 psig); 1.76 kg/cm (25 psig) optimum
  • Gas and Ambient Temp: 5 to 50°C (41 to 122°F)
  • Leak Integrity: 1 x 10-7 cc/sec of He max to outside environment
  • Materials in Fluid Contact:
    • Aluminum Models: Anodized aluminum, 316 SS, brass and FKM O-rings
    • Stainless Steel Models: 316 SS and FKM O-rings
  • Output Signal:
    • Linear 0 to 5 Vdc: 1000 Ω minimum load
    • 4 to 20 mA: 50 to 500 Ω loop resistance, ±20 mV max noise
  • Transducer Power: 12 to 26 Vdc @ 800 mA
  • Turndown Ratio: 40:1
  • Shipping Weight: 1.8 kg (4 lb)
  • Compliance: EN55011 class 1, class B; EN50082-1
  • Attitude Sensitivity: With horizontal flow path, no shift in calibration up to +20 degree pitch
  • Gas Relative Humidity: 0 to 70% RH
  • Min Pressure Drop