SieversModel M5310 C -On-line TOC Analyzer

SHARE
Dual-stream, continuous monitoring designed to meet regulations and optimize processes. The Sievers M5310 C On-Line Total Organic Carbon (TOC) Analyzer offers both continuous and discrete grab samples with dual-stream capabilities. Online capabilities allow for real-time monitoring, while the Integrated On-Line Sampling System (IOS System™) allows for the introduction of external samples. Offering an operating range of 4 ppb to 50 ppm, this analyzer is engineered to support regulatory compliance and processes optimization goals.
Most popular related searches

The dual-stream and continuous monitoring capabilities of the Sievers M5310 C On-Line enable efficient, real-time monitoring of two independent sample sources. The analyzer provides automated calculations of percent TOC removal for influent and effluent streams. Make real-time decisions with the Sievers M5310 C On-Line TOC Analyzer. 

  • Provides readings in only two minutes
  • Automated calibration, verification, and data analysis enable unattended operation
  • Automatically calculates percent TOC removal for influent and effluent streams or samples
  • The Autoreagent feature automatically establishes optimal flow rates for each sample
  • The grab-mode, alphanumeric labeling feature allows individual labeling of each grab sample for easy recall later on
  • Two-stream sample line configuration enables efficient sampling of two independent sample sources, without the need to alter the initial inlet setup
  • Pre-calibrated for simple installation 
  • Minimal maintenance — typically just a few hours per year 
  • Self-contained with small footprint and internal reagent packs — no external chemicals, catalysts, or gas supplies required 
  • Easy data communications — export or collect data via USB, 4-20 mA, or Modbus TCP/IP outputs
  • Integrated On-Line Sampler (IOS) system allows testing of external samples and standards without removing the analyzer from the continuous water source

Operating Specifications*

  • Range 4 ppb to 50 ppm
  • Precision < 1% RSD
  • Accuracy ± 2% or ± 0.5 ppb, whichever is greater
  • Display Readout 3 significant digits
  • Sample Type On-line continuous or discrete grab sample, timed on-line, TOC removal on-line, TOC removal grab
  • Calibration Typically stable for 12 months
  • Analysis Time 2 minutes
  • Sample Temperature** 5–95 °C (41–203 °F)
  • Ambient Temperature 4–40 °C (41–104 °F)
  • Sample Pressure Up to 100 psig
  • Required Sample Line Flow Rate >50 mL/min (for on-line mode)
  • Instrument Sample Flow Rate 0.5 mL/min

Analyzer Specifications

  • Inlets One stream, or two-stream inlet (option)
  • Outputs 4-20 mA outputs (3); alarm outputs (4); binary input (1);  USB device port (1), USB host ports (2); Ethernet
  • Display 7” WVGA 800x480 pixel, Color  LCD w/ touch-screen
  • Power 100 – 240 V~, 50 – 60 Hz, 100 VA
  • Dimensions H: 54.9 cm (21.6 in); W: 45 cm (17.7 in); D: 26.5 cm (10.4 in)
  • Weight 16.2 kg (35.6 lb)
  • Safety Certifications ETL, CE see Declaration of Conformity
  • Enclosure Rating IP-45

Environment

  • Maximum Relative Humidity 0 - 95%, non-condensing
  • Maximum Altitude 3,000 m (9,800 ft)
  • Pollution Degree 2
  • Prefilter Kits: provide sample conditioning to ensure optimal instrument performance. Three application-specific kits are available:  normal use, heavy use, and treated drinking water