Tethys Instruments SAS - HORIBA Advanced Techno

TethysModel UV300-5D -Online Analyser for Drinking Water

SHARE

The UV300-5D is a pre-defined configuration of the UV300 analyser dedicated to drinking water monitoring.

Most popular related searches

It offers the possibility to monitor simultaneously FIVE parameters :
(1) Total residual chlorine    (2) Turbidity    (3) pH    (4) Conductivity    (5) Temperature
All this is done by a single unit that allows high performance measurements at a cost effective price. An automatic cleaning system maintains clean the flow cell and also make an auto-zero that guaranties the absence of long term drift.

Colorimetric DPD Method : A Reliable Method for Total Residual Chlorine

  • The US-EPA 330.5 (or standard methods 4500-Cl-G) DPD method is internationally recognized for its good accuracy and stability, while specific electrodes for chlorine may drift or show a sensitivity to the sample flow.
  • Other oxidising agents are also taken in account by the DPD method, that makes it the best choice method.
  • The colorimetric module has been specially developed to reach a very small volume flow cell (7ml) that reduces the quantity of reagent to preserve the environment and reduce the operating cost.
  • A multi-wavelength LED source assumes a colour and turbidity compensation with an unlimited life time. Large bore tubing avoid clogging as accumulated sediments might be a problem even on drinking water for long term operation.
Turbidity by Laser Diode (Nephelometric)

The turbidity module is based on a high stability laser diode to reach low level of detection as required for drinking water. The automatic cleaning system allows to perform an auto-zero once per day to guaranty the absence of drift on long term. It avoid also the accumulation of sediments that may perturb the measurement. The laser diode could be either a red visible model to ease the maintenance and troubleshooting, or a infra-red model to follow the ISO 7027 nephelometric method at 860 nm.

Patented Flow cell
  • The patented flow cell allows very high level of suspended solid without clogging for all the optical measurements making it suitable for industrial and municipal waste water application. The turbidity is automatically compensated by a dual-wavelength method.
  • Suitable for sample with high chloride content
  • Wetted parts of flow cell make it suitable for most corrosive sample
Multiplexing system
When different streams need to be analysed, for example inlet and outlet of a plant, an optional multiplexing system delivers relay contacts to control external electric-valves or external pumps. Up to 6 different streams can be selected
The measuring channels can be either duplicated (each one having its own 4-20mA output or MODBUS register), or measured sequentially to fit with the maximum of 16 measuring channels (a MODBUS register indicates which stream is currently being measured).
Communication The RS232 port supports the MODBUS protocol to transmit each Standard channel value to a SCADA system.
Additional parameters are available like status code, error code, calibration values and pumps run time. Basic 4-20 mA output modules can be plugged on the main board for each measuring channel. A USB port enables to download on any USB key the last 5000 recorded measurements as well as a diagnostic file containing the configuration and useful information for remote troubleshooting. The new web interface makes possible to drive remotely the analyser from any computer, tablet or i-phone with a web browser. For this, an optional Wi-Fi or Ethernet module is added inside the analyser to connect it to an existing network with an internet gateway. The recorded measurements file can be imported to Excel for graphs or other treatments. The software of the analyser can be upgraded by connecting a USB key.