OxyCycler - Model A84XOV - Dynamic O2 Controller for Animal Modeling
From Animal Products
The OxyCycler A84XOV is a powerful research tool for scientists who do oxygen sensitive work. It makes complex oxygen profile control easy.
-
Most popular related searches
Products Details
Dynamic O2 Controller for Animal Modeling
- Reproducible, dynamic O2 profiles
- Works in any semi sealable container!
- Runs four independent protocols simultaneously
- Works with BioSpherix animal modeling equipment for in vivo research
- User-friendly PC interface
- Data logging
- Intermittent Hypoxia In Vivo
- Hypoxia/Hyperoxia Modeling In Vivo
FULL RANGE DYNAMIC PROFILES
Control oxygen profiles with multiple setpoints anywhere from 0.1-99.9% oxygen.
Hold any setpoint for any length of time. Rate of change between any two setpoints is fully adjustable. Cycle any profile any fixed number of times, or continuously with this animal modeling oxygen controller.
WORKS IN ANY CHAMBER
The OxyCycler A84XOV works exceptionally well with BioSpherix A-Chambers. However, it can work with practically any semi-sealable enclosure. Large or small. Square or round. Flexible or rigid. Manufactured or custom made. Most chambers can be fitted in minutes.
MULTI-CHAMBER IS PRODUCTIVE
Controls in four chambers at once. Operate all chambers simultaneously, or each independent of the others. Control different profiles, or same profile in every chamber. Start all at the same time, or each at different times.
CONTROL IS EFFICIENT
Nitrogen is infused to lower oxygen. Oxygen is infused to raise it. Feedback from oxygen sensor inside chamber tells the OxyCycler A84XOV exactly when and exactly how much gas to infuse. No gas is ever wasted! Any disturbances are immediately detected and corrected.
OPERATION IS SIMPLE
Once the animal modeling oxygen controller is installed and configured, it’s easy. Occasionally check calibration and supply gas, otherwise it’s all automatic! Can work continuously year round, or on occasion as needed.
Specifications
- ELECTRICAL POWER: 12 VDC @ 6.66A
- CONTROL RANGE: 0.1-99.9% oxygen
- ACCURACY: ±1%
- RESOLUTION: 0.1%
- GAS SOURCE: Compressed gas tanks, liquid carboys (from headspace), or generators.
- GAS SUPPLY: Pressurized O2 and N2
- GAS SUPPLY LINE: 1/4 inch I.D. hose pressure rated at 40 PSIG.
- GAS SUPPLY LINE PRESSURE: 0-40 PSIG
- GAS INFUSION RATE: 1-150 S.C.F.H. each control gas each chamber.
- GAS SUPPLY HOSE FITTINGS: 1/4 inch hose barb.
- UMBILICAL LENGTH: 12 feet (custom lengths available).
- ACTUATOR POD SIZE: 7”H, 4.375”W, 4.5”D
- ALARM OUTPUT: Visible flashing indicator. PC adds audible and more visible indicators.
- ALARM MODES: Process high, process low, deviation high, deviation low, deviation band.
- WEIGHT: 31 lbs. (Controller Only)
- DIMENSIONS: 9”H x 21.2”W x 18.5”D
Sensor Operational Parameters
- HOST CHAMBER TEMPERATURE: 0-40°C (depending on sensor).
- HOST CHAMBER CO2: 0-20% (depending on sensor).
- HOST CHAMBER HUMIDITY: 5-95% non-condensing
Customer reviews