Pacon - Model 5000 -Online Water Hardness Analyzer
PACON 5000 is a compact, easy-to-operate and high accuracy water qualityanalyzer provided by JENSPRIMA for automatic on-line inspection of water quality residual hardness and quality control of water softening processes. This system controls the selectable limits based on the titration colorimetric principle, providing accurate measurement readings by extinction, and a variety of functions to ensure the reliability of real-time operation.Low maintenance and low reagent consumption, suitable for long time continuous operation, maintenance-free, especially suitable for the boiler room.
- Process Water
- Water Treatment
- Drinking Water
- Cooling Water
Fully-automatic Measurement
- Full-automatically measure the total water hardness with different ranges according to the selected reagent.
- The analysis process is more efficient than manual measurements and also more reliable than other indirect
- measurement methods, such as ion-selective electrode.
Intelligent & Accurate
- GB measurement method - titration colorimetric method, the instrument does not require calibration. The
- integrated measurement technique and the two-stage analysis process can identify external
- measurement effect,such as contamination of the cell, turbidity of the water sample and external light, and to
- eliminate these effects in the measurement.
Optional Measurement Interval
- Optional measurement interval: 5-360min
- PACON 5000 can also control the start-up measurement of the instrument via an external switching signal.
Compact Design
- With size 300X300X200mm, it is suitale for direct wall mounting or install on the bracket.
Least Maintenance Workload
- The measuring slot must be cleaned according to the set measuring interval or the measuring frequency.
- It is recommended to replace the spare parts every year. The spare parts including: peristaltic pump head,
- reagent connection tube and seal, Order No. 50-5000-10. No additional tools are required for maintenance and can be easily executed.
Automatic Cleaning
- Each analysis will automatically perform Rinsing and Cleaning, ensuring measurement accuracy, repeatability and reduced on-site maintenance.
Low Reagent Consumption
It is very easy to replace the reagent bottle, 500ml reagent can measure 5000-10000 times. Reagent is valid for 2 years.
LCD Display
Multi-language graphic backlit LCD display, showing measured values, reagent remaining, alarm values and relay status.
SD Card Storage
- 2G data memory card, can be directly connected to the computer, to access to historical data and system failure information in excel format
Output & Input
- 0 / 4-20mA & CAN bus
- Insufficient reagent
- Alarm signal
- External start analysis signal
General Specification
- Measurement method: Titration method with colour change
- Environmental temperature:5 - 45℃
- Measuring water temp.:5 - 40℃
- Water inlet pressure: ca. 0.5-5bar, recommended 1-2bar
- Inlet / Outlet connect: 6mm hose
- Requirements of the water quality:
- clear,colourless,no solid particles,
- without gas bubbles; pH: 4 - 10.5 ;
- Iron: < 3 ppm ; Copper: < 0.2 ppm ;
- Al: < 0.1 ppm ; Mn: < 0.2 ppm;
- Humidity: 20 - 90% RH
- Supply Voltage: 85 - 265VAC,47-63Hz
- Power consumption: 25VA(in operation)
- Dimensions: 300 x 300 x 200 mm (W x L x H)
- Weight: ca.5Kg
- Protection class:IP65
- Installation: Wall mounting in closed rooms
Analysis Specification
- Measurement range :The measurement range is defined by the reagent used. (see reagent table)
- Measurement duration: ca. 3 minutes depending on the hardness of the water
- Measurement accuracy: ±5% of the upper value of the respective reagent
- Repetition accuracy: ±2.5% of the upper value of the respective reagent
- Analysis cycle: Measuring interval(5 - 360min) /External signal
- Flush time:15 - 1800s
- Display: Multi-coloured and multi-lingual graphic display
- Unit: mmol/l、 dH、mg/l、ppm CaCO3
- Analogue output: 0/4~20mA,max. 750Ω
- Interface: CAN bus
- Relay output: 4x potential free output NC/NO, 250VAC/DC, 4A
- Input: IN1: Start analysis, IN2:Cancel error