PoliSimeter™ series of electronic personal dosimeters (EPDs) is the next generation of the well-known PM1610 series by Polimaster, which was a trusted solution on the market for nearly 15 years, with over 20,000 dosimeters sold. Building on the legacy of the PM1610, the PoliSimeter offers enhanced capabilities for the extended measurement of personal dose Hp(10) and dose rate Ḣp(10) of both continuous and pulsed X-ray and gamma radiation. The PoliSimeter is designed to meet the latest industry requirements and features a USB-C port for convenient connectivity and data transfer. PoliSimeter ERB model has an extended range of dose measurement of up to 20 Sv for use in high-radiation environments, providing critical safety and monitoring functionality. Instead of a rechargeable battery, this model is powered by an AAA battery, which is easy to replace, affordable and safe to handle.
Most popular related searches

Instruments provide measurement of dose equivalent and dose equivalent rate of continuous and pulsed X-ray and gamma radiation, measurements can be displayed in Sv or rem. Two independent alarm thresholds for dose and dose rate can be set. Instruments have wide energy range from 20 keV to10 MeV, compact size and light weight, audible, visual and vibration alarms. Dosimeters are equipped with AAA (LR03) battery, so can be used immediately after installing the battery.

Instruments can be integrated into Personal Dose Tracker System. PDT System is intended for continuous account for external radiation exposure, maintenance of dose history database and calculation of the personnel accumulated doses (individual as well as group doses) during the set period of time, automated preset threshold (individual as well as group thresholds) alarm signal. Software enables information transfer by USB-protocol to a personal computer to analyze, process, and generate corresponding databases in a control center or an expert center.

  • Detector: Geiger-Muller tube
  • Dose equivalent rate (DER) measurement range: 0.1 μSv/h to 10.0 Sv/h
  • DE indication range: 0.001 μSv to 24.0 Sv
  • DE measurement accuracy: ± 20 %
  • DER measurement accuracy: ±(10 + K1/? + K2*?) %, where H - DER value in mSv/h, K1 − coefficient 0.0015 mSv/h, K2 − coefficient 0.0015 (mSv/h)-1:
  • Energy range: 0.02 to 10.0 MeV
  • DER indication range: 0.01 μSv/h to 12.0 Sv/h
  • Energy response relative to 0.662 MeV:
    • from 20 to 33 keV: - 60 %
    • from 33 to 48 keV: - 40 %
    • from 48 keV to 3 MeV: ± 30 %
    • from 3 to 10 MeV: ± 50 %
  • Thresholds: DER thresholds setting range - 0.01 μSv/h to 10.0 Sv/h; DE thresholds setting range: 1.0 μSv to 20.0 Sv
  • Alarm Type: Audio, visual, vibration
  • Drop test on concrete floor: 1.5 m onto hard wooden surface
  • Ingress protection: IP65
  • Power supply: battery 1.5 V, AA? (LR03), no less than 1100 mA/h; external power supply
  • Battery Lifetime:
  • average DER not more than 0.3 μSv/h, backlight, sound, light and vibration alarms activated for not more than 20 s/24 hours: no less than 20 days
  • Weight: 0.09 kg
  • Operating temperature: -20 to + 50 °?
  • Atmospheric pressure: 84 to 106.7 kPa
  • Dimensions: 71 ? 59 ? 20 mm
  • Humidity: 98 % at 35 °?
  • Mean service lifetime: 10 years
  • Data exchange: USB
  • DE measurement range:
    • continuous photon radiation: 0.05 μSv to 20.0 Sv
    • pulsed photon radiation (pulse duration not less than 1 ms): 10 μSv to 20.0 Sv