Streamline Measurement Ltd.

ProloggerModel 7001D -Data Loggers

SHARE

The Prologger 7001D-1 sets new standards in accuracy and input range. It is packaged in the robust enclosure so familiar to Starlog users with twice the accuracy, eight times the resolution, and sixteen times the dynamic range of its predecessor. All Prologger analogue and digital inputs are processed with 16 bit resolution.

Most popular related searches

The sixteen inputs now support the following ranges:

  • ±5.000V (155uV/bit resolution)
  • ±500mV (15.5uV/bit resolution)
  • ±50mV (1.55uV/bit resolution)
  • ±5mV (155nV/bit resolution)

The superior accuracy of the design means that input voltages will be converted to better than 0.05% of full scale over the full operating temperature range, and 0.1% in the 5mV range.

The Prologger’s increased memory capacity means you can acquire more data or increase the period between downloads. The unit also includes all the familiar Starlog features such as SDI-12 instrument support, modem command/dial-out support, universal battery pack, continuous power source, scheme control of power supplies, and field upgradable control firmware.

The Prologger can also be used with the Neon Internet logging system allowing users to view up to th eminute data directly on their PC, Laptop, Tablet or Smart Phone.

  • Very accurate - 16 bit resolution
  • Wide input signal range
  • Power input from solar panel, DC power supply or 12V battery
  • Large 1Mbyte memory
  • 64 character display
  • Long battery life
  • SDI-12 support
  • Cost effective
  • Compatible with Neon Internet system
  • Multiple input ranges
  • External control outputs
  • Material: Grey, high impact, rigid PVC
  • Size: 211mm x 108mm x 81mm (HxWxD)
  • Weight: 2kg (including battery)
  • Operating temperature: –20°C to 60°C. Not affected by humidity
  • Scan rate: 0.125 seconds to 5 minutes - programmable
  • Log interval: 0.125 seconds to 1 week - programmable
  • Memory: Low power CMOS RAM 1Mb standard
  • Time clock: Crystal regulated, ±10 seconds per month
  • Analog inputs:
    • 16 channels, 16 bit resolution on all channels.
    • Unipolar or bipolar, differential or single-ended
    • Voltage input in four programmable ranges
    • 5.00V to +5.00V, 155u V/bit resolution
    • 500mV to +500mV, 15.5uV/bit resolution
    • 50mV to +50mV, 1.55uV/bit resolution
    • 5mV to +5mV, 155nV/bit resolution
  • Counters:
    • 4 channels, 16 bit resolution.
    • DC to 20kHz potential free contacts
    • Accepts 0 to 12V DC digital inputs
    • (0 to 1 threshold = 5V)
  • STARBUS:
    • 2 x high speed serial lines with eight
    • channels on each. 16 bit, bi-directional,
    • synchronous data and clock
  • SDI-12:
    • Up to ten SDI-12 sensors are supported.
    • SDI-12 Standard V1.3
  • Controls:
    • 2 channels, 1 CMOS output. 1 uncommitted
    • open collector output
  • Computer Input/Output: Full duplex serial RS232C.
  • Baud rates: 300-76800
  • CPU: 80C31 microcontroller, 14.7456 MHz
  • Battery life: Alkaline 1 year (typical)
  • Flat battery shutdown: 5.6 Volts
  • Power:
    • 5V DC reg. 100mA, 6.5V unreg. 1mA
    • cont.,10V DC reg. 100 mA prog. duty cycle
    • (PDC), 12V DC unreg. 200 mA (PDC), -12V
    • DC unreg., 50 mA (PDC)