Applications: Measurements at two Sample Points. Emission Monitoring. Optimizing Processes. Combustion Control. Threshold Limit Values (TLV) - Measurements.

Most popular related searches
  • Heated Analytical Compartment (up to 160 °C / 320 °F)
  • 2 Sample Inputs, Manual or Auto-Control
  • Continuous  / Discontinuous Measurement Mode
  • Flame Out Indication
  • Auto-Range / Manual Range
  • Automatic Hydrogen Cut-Off
  • Self-Monitoring Capabilities (Temperatures, Pressures, Flame)
  • Option: Sample Lines
  • Option: Filters
  • Option: Internal Zero-Gas and Combustion-Air Supply
  • Option: USB-Interface

(Reference Quantity : Full Range Deflection = FRD)

  • Ranges:
    • 0-10 ppm Propane
    • 0-100 ppm Propane
    • 0-1.000 ppm Propane
    • 0-10.000 ppm Propane
    • 0-100.000 ppm Propane
  • Linearity
  • Sample Input Select
  • +/- 2 % FRD (up to Range 4)
  • Manual or Automatic, Time Base 1-99 Minutes
  • Oxygen Cross Sensitivity: < +/- 2 % FRD
  • Availability: > 95 %
  • Limit of Detection: < 2 % FRD
  • Zero Drift: < 2 % FRD/week
  • Span Drift: < 4 % FRD/week
  • Response Time: approx. 1-2 s (without sample line)
  • Sample Flow Rate: approx. 220 l/h
  • Signal Output: 0-10 V or 0/4-20 mA
  • Status Contacts:
    • Pre (Concentration-) Alarm
    • Main (Concentration-) Alarm
    • Range Identification
    • Flame Out Indication
    • Status Identification (Zero-Gas/Span- Gas/Calibration)
    • Alarms (Pressure, Temperature, Flame)
  • Dimensions: 19" x 267 x 500 mm (WxHxD)
  • Weight: approx. 26,5 kg
  • Ambient Temperature: + 5 ... + 35 °C
  • Main Power:
  • 230 V / 50 Hz or 115 V / 60(50) Hz / 800 Watts