The REFEX Reference System has a higher impedance than conventional porous junctions therefore the impedance setting needs to be changed on the Yokogawa instrument.
- Enter the SETUP
- Select COMISSIONING - ENTER
- Select MEASUREMENT SETUP - ENTER
- Select IMPEDANCE SETTINGS - ENTER
- Switch setting from LOW 10 HIGH IMPEDANCE on INPUT 2