Breas Medical AB
  1. Companies
  2. Breas Medical AB
  3. Products
  4. Clearway - Model 2 - Second Generation ...

ClearwayModel 2 -Second Generation MI-E Device

SHARE

Introducing the new Clearway 2 with TreatRepeat® The second-generation MI-E device by Breas incorporates the latest technology to bring effective and accurate treatment for patients requiring airway clearance and cough assistance. TreatRepeat function allows clinicians to manually deliver different numbers of insufflations and exsufflations and then save the treatment they have just delivered to the device to be repeated automatically. Helping to improve accuracy and speed of titration and set-up.

Most popular related searches

ADVANCED TITRATION TECHNOLOGY

  • Unique to the Clearway 2 – TreatRepeat®
  • Intuitive touchscreen user interface
  • Hand-held remote manual switch with start/stop function

MULTIPLE MODE OPTIONS

  • Programmable automatic modes
  • Dedicated IPPB mode
  • Manual mode

ENHANCED PATIENT SYNCHRONY

  • TreatRepeat® to help improve accuracy and speed of titration and set-up
  • iTrigger® − built on the Breas patented eSync algorithm
  • Synchrony Beep® to aid patient-device synchrony
Settings / Performance
  • Modes : Manual, Basic Auto, Program Mode, IPPB, NIV
  • Device Modes : Home, Clinical (Locked/Unlocked)
  • Profiles : 4 settable profiles
  • Insufflation Pressure : 3 cm H2O to 70 cm H2O
  • Exsufflation Pressure : −3 cm H2O to −70 cm H2O
  • Insufflation Time : 0.5 to 5 seconds
  • Exsufflation Time : 0.5 to 5 seconds
  • Pause Time : 0.5 to 5 seconds
  • Flow Options (MI-E) : 1−10 (10−100% of Ti max 4 seconds)
  • Flow (IPPB) : 20 L/min to 120 L/min
  • Inspiratory Trigger : Off or 1 to 10
  • Oscillations Frequency : Insufflations and or Exsufflations, Off – 1 – 20 Hz
  • Oscillations Amplitude : 1 cm H2O to 10 cm H2O
Monitoring
  • Displayed Data : Cough Peak Flow, Insufflation Volume, Pressure, Flow, SpO2, Heart Rate Sessions Used, Average Session Duration
  • Waveforms : Pressure
  • Trend and Statistical Data via PC Software : Cough Peak Flow, Insufflation Volume, Use, SpO2, Heart Rate
Power Supplies
  • Mains Supply : 100 to 240 V AC
  • Internal Battery : >30 min in MI-E modes (based on settings of +40/ -45cmH2O, Ti/Te/P 2 seconds)
Main Alarms
  • Alarms : Circuit Disconnection, Running on Battery Power, External Battery Fault, High Pressure Alarm, Internal Over Temperature Alarm, Low Battery, Low Memory Battery, Machine Fault, Low SPO2 Reading, High SO2 SpO2 Sensor Disconnect, High/Low Heart Rate, Power Fail
Dimensions
  • W x H x D : 285 x 285 x 195 mm
  • Weight : 3.85 kg
  • Noise Level (at 10 cm H2O Constant Pressure) : Sound Level at 70 cm H2O in Manual Mode Less Than 74 dB(A) Measured at 1 meter
Connectivity and Software
  • Built-in Connectivity : Ethernet, USB, SD card
  • Compatible Software : Prepared for Breas Cloud Service
  • Plug-and-Play Accessories : SpO2, Hand Held Remote Start/Stop and Manual Switch, Manual Foot Pedal Switch
Other
  • Circuits : Disposable circuit with bacterial filter no leak port for MI-E modes and disposable circuit with bacterial filter and leak port for IPPB and NIV modes