Seegene STARletOne-Step Process from Nucleic Acid Extraction

SHARE

One-step process from nucleic acid extraction to PCR setup. Easy NA extraction and PCR setup for multiple specimens. Seegene STARlet IVD is an easy-to-use liquid handling workstation from primary sample tube to nucleic acid (NA) extraction and PCR setup. It provides convenient process of your lab works by minimizing hands-on time and maximizing assay reliability.

Most popular related searches

Flexible solution for extraction or PCR setup selectively

  • One step process from extraction to PCR setup
  • Extraction-only operation
    • PCR setup-dedicated operation
  • Can be used with other NA extraction instruments
    • PCR setup-dedicated operation

Direct loading of primary sample tubes

  • 1.5 ml tube
  • 12 mm tube
  • 16 mm tube
  • Thinprep

A single set of reagent for extraction of bacterial, viral, genomic, parasitic, fungal DNA and/or RNA from various types of specimens
Simultaneous PCR setup up to 4 assays together with in-house PCR assays customized in your laboratory

  • Power input: 115-230V, 50-60 Hz
  • Power consumption: Maximum 600 W
  • Dimensions: 1124 (W)x795 (D)x903 (H) mm
  • Weight: 140 Kg
  • Sample capacity: 1-96 samples
  • TAT: 155 min for whole proce
  • Pipetting channel: 8 channels
  • Dispensing precision
  • (when using 300? tip): 10? : 2%, 50? : 0.75%, 200? : 0.75
  • Dispensing precision
  • (when using 1000? tip): 10? : 3.5%, 100? : 0.75%, 1000? : 0.75%
  • Positional accuracy: 0.1 mm on X-Y-Z
  • Ordering Information: 67930-03