Sievers - Model M5310 C -Laboratory TOC Analyzers
Sievers M5310 C Laboratory Total Organic Carbon (TOC) Analyzers are designed for the drinking water industry and environmental water quality monitoring. The analyzer is easy to use and maintain. It offers the ability to be used for single grab samples or with the Sievers Autosampler. The Sievers M5310 C Lab is engineered to support disinfectants and disinfection byproducts rule (DBPR) compliance, with an operating range of 4 ppb to 50 ppm.
Quick to set up and easy to maintain, the Sievers M5310 C Lab provides automated calculations of percent TOC removal for influent and effluent streams. Meet regulatory goals and make data-driven decisions with the Sievers M5310 C Lab.
- Provides readings in only two minutes
- Automated calibration, verification, and data analysis enable unattended operation
- Automatically calculates percent TOC removal for influent and effluent streams or samples
- Can be used with the high-capacity Sievers Autosampler, featuring random access capability and up to 120 sample positions.
- The Autoreagent feature automatically establishes optimal flow rates for each sample
- Pre-calibrated for simple installation
- Minimal maintenance — typically just a few hours per year
- Self-contained with small footprint and internal reagent packs — no external chemicals, catalysts, or gas supplies required
- Easy data communications — export or collect data via USB or Modbus TCP/IP outputs
Operating Specifications*
- Range 4 ppb to 50 ppm
- Precision < 1% RSD
- Accuracy ± 2% or ± 0.5 ppb, whichever is greater
- Display Readout 3 significant digits
- Sample Type Autosampler or discrete grab sample
- Calibration Typically stable for 12 months
- Analysis Time 2 minutes
- Sample Temperature 5–95 °C (41–203 °F)
- Ambient Temperature 5–40 °C (41–104 °F)
- Instrument Sample
- Flow Rate 0.5 mL/min
Instrument Specifications
- Outputs USB device port (1), USB host ports (3); Ethernet
- Display 7” WVGA 800x480 pixel, Color LCD w/ touch-screen
- Power 100 – 240 V~, 50 – 60 Hz, 100 VA
- Dimensions
- H: 42.2 cm (16.6 in.); W: 24.6 cm (9.7 in); D: 40.0 cm (15.8 in)
- Weight 9.7 kg (21.4 lb)
- Safety Certifications ETL, CE see Declaration of Conformity
Environment
- Maximum Relative Humidity 0 - 95%, non-condensing
- Maximum Altitude 3,000 m (9,800 ft)
- Pollution Degree 2
- Sievers Autosampler: enables 24+ hours of unattended sample analysis (63 or 120 sample position capacity)
-
DataPro2 Software: integrates the Autosampler with the M5310 C Lab, providing timesaving features to maximize productivity and ensure easy TOC data management
-
Autosampler rinse station: flushes the sample inlet system and needle with a flowing water source, maximizing analytical performance of samples of varying concentrations