Signal Conditioner Optimized for Measuring Physiological† Pressure. The LifeSens is a compact signal conditioner specifically designed for measuring physiological† pressure. When combined with OpSens’ OPP-M fiber optic sensors, the LifeSens delivers outstanding pressure measurements with all the features required for life science applications.
The LifeSens benefits from OpSens’ White Light Polarization Interferometry (WLPI) technology (Patent # 7,259,862), providing the most accurate and reliable absolute Fabry-Perot cavity length measurements.
The LifeSens delivers pressure measurements at a rate of 250 Hz. An internal manometer automatically corrects for atmospheric pressure change. The system accepts the OpSens’ SCPROM connector that stores OPP-M sensor gauge factor. Upon mating the sensor, the gauge factor is seamlessly transmitted to the LifeSens conditioner.
The LifeSens is equipped with a large visible LCD display and can be battery operated. It comes with a standard ±5 V output and a RS-232 communication port for real-time data acquisition. The LifeSens can be controlled directly using the front-panel keypad or remotely using the standard RS-232 interface. A rugged casing with a removable rubber boot provides good mechanical protection against intensive handling in tough environments.
Other relevant literature: IMP0006 OPP-M.
- Compact and rugged design
- 250Hz sampling rate
- High resolution and precision
- Compatible with SCPROM Connector
- ±5 V and RS-232 output interfaces
- Large LCD display
- 9V battery or wall-transformer operated
- Automatic atmospheric pressure correction
- OEM version available
- Physiological pressure monitoring:
– Intra vascular blood pressure
– Urodynamic
– Intra cranial pressure
– Intra uterine pressure
– Intra ocular
– Cardiac assist applications - Customer specific catheter application
- Guidewire design (with 0.25 mm OD transducer)
- Pressure measurement under electro-surgery,
- MRI and MW/RF related practices
- Number of channels One
- Compatibility Opsens’ OPP-M fiber optic pressure sensor (other sensors upon request)
- Full scale* -50 mmHg to 300 mmHg relative to atm. (wider range also available)
- Resolution* 0.5 mmHg (no averaging)
- Precision* 1% FS or 1 mmHg (whichever is greater)
- Sampling rate 250 Hz standard
- Connector compatibility SC connector (Opsens’ SCPROM connector compatible)
- Internal manometer Included for automatic atmospheric pressure correction
- Output interface ±5 V (1V/100mmHg) and RS-232 standard
- Input power and consumption 9 to 24 VDC – 1.8 W (wall-transformer adapter included)
- Dimensions (without rubber boot protection) 45 mm (H) x 105 mm (W) x 165 mm (L)
- Storage temperature -40 °C to 70 °C
- Operating temperature 0 °C to 45 °C
- Humidity 95 % non condensing
- Light source life span 40 000 hours MTBF