Spectrum Techniques, LLC
- Home
- Companies
- Spectrum Techniques, LLC
- Products
- Spectrum - Model GM35 - End-Window GM ...
Spectrum - Model GM35 -End-Window GM Detector
The GM35, end-window, Geiger-Mueller detector will detect alpha, beta, and gamma, and x-ray radiation (or combinations thereof).
Most popular related searches
- Maximium Starting Voltage : 850 volts
- Recommended Operating Voltage : 900 volts
- Operating Voltage : 850-1000 volts
- Maximum Dead Time : 150 µs
- Maximum Plateau Slope : 10% / 100 volts
- Minimum Anode Resistor : 1 MΩ
- Recommended Anode Resistor : 3.3 MΩ
- Gamma Sensitivity Cs-137 : 40 cps per mR/hr
- Tube Capacitance : 5 pF
- Nominal Active Length: 2.25 inches (57.2 mm)
- Window Material: Mica
- Window Thickness: 1.8 to 2.2 mg/cm2
- Active Diameter: 1.13 inches (28.7 mm)
- Cathode Material: Cr – Fe
- Cathode Wall Thickness: 0.06 inches (1.52 mm)
- Operating Temperature Range: -20 to 50°C
- Fill Gas: Ne/Ar + Halogen
- Type of Connector: BNC
- Approximate Weight: 150 grams
- Maximum Altitude: 8000 ft