- Home
- Companies
- SUEZ Water Technologies and Solutions
- Products
- SUEZ - Model M9 - Sievers On-Line TOC ...
SUEZ - Model M9 -Sievers On-Line TOC Analyzer
The Sievers M9 On-Line Analyzer offers excellent low-level sensitivity for ultrapure water analysis and high-level capability for cleaning validation or other challenging water samples. It’s proprietary Integrated On-Line Sampling (iOS) System enables easy introduction of external standards and samples without removing the instrument from the continuous sample source. With Sievers unique technology, the M9 On-Line is immune to false positive and false negatives that trouble non-membrane based TOC instruments.
Designed for continuous monitoring of waters from ultrapure to potable, the M9 On-Line provides unparalleled accuracy, precision, and speed across a dynamic operating range of 0.03 ppb to 50 ppm. It supports compliance with pharmacopeia water testing or can be used for cleaning validation and other online applications. In Turbo mode, the M9 provides analysis every four-seconds, providing up-to-the-second results for better control. Sievers M9 Analyzers meet or exceed the most common pharmaceutical regulatory requirements for TOC analysis and data integrity.
- Two-minute sample analysis time
- Continuous monitoring with the ability to easily introduce discrete grab samples and standards
- Wall-mounted with an IP-45 rated enclosure to withstand demanding environments
- Automated operations, such as calibration and data analysis
- Autoreagent feature automatically establishes optimal flow rates for each sample
- Minimal preventive maintenance (typically just a few hours per year)
- 12-month calibration stability
- Optional Turbo mode for four-second analysis
Operating Specifications*
- Range 0.03 ppb to 50 ppm
- Precision < 1% RSD
- Accuracy ± 2% or ± 0.5 ppb, whichever is greater
- Display Readout 3 significant digits
- Sample Type On-line continuous or discrete grab sample
- Calibration Typically stable for 12 months
- Analysis Time 2 minutes (4 seconds with the optional Turbo mode)
- Sample Temperature** 5–95 °C (41–203 °F)
- Ambient Temperature 5–40 °C (41–104 °F)
- Sample Pressure 100 psig
- On-Line Flow Rate >50 mL/min (for on-line mode)
- Instrument Sample
- Flow Rate 0.5 mL/min
Analyzer Specifications
- Outputs 4-20 mA outputs (3); alarm outputs (4); binary input (1); USB device port (1); USB host ports (2); Modbus TCP/IP
- Display 7" WVGA 800X480 pixel, Color LCD with touch-screen
- Power 100 - 240 V~, 50 -60 Hz, 100 VA
- Fuses Replace with same type and size fuse: T 1.6 A 250 VAC Fuse (Slow Blow), size 5 X 20 mm appliance inlet
- Dimensions H: 54.9 cm (21.6 in); W: 45.0 cm (17.7 in); D: 26.5 cm (10.4 in)
- Weight 15.8 kg (34.9 lb)
- Enclosure Rating IP-45
- Safety Certifications ETL, CE (see Declaration of Conformity)
Environment
- Maximum Relative Humidity 0 - 95%, non-condensing
- Maximum Altitude 3,000 m (9.800 ft)
- Pollution Degree 2
Operating Specifications*
- Range 0.20 ppb to 10 ppm
- Precision 2% RSD
- Accuracy ± 10% or ±10 ppb whichever is greater
- Display Readout 3 significant digits
- Sample Type On-line, Grab
- Calibration Typically stable for 12 months
- Analysis Time 4 seconds
- Response Time 3.25 minutes
- Sample Temperature 5–95 °C (41–203 °F)
- Ambient Temperature 10–40 °C (50–104 °F)
- Sample Pressure 100 psi
- Sample Flow Rate 30–300 mL/min (for on-line mode)
- Instrument Sample Flow Rate: 1.1 mL/min
Analyzer Specifications
- Outputs 4-20 mA outputs (3); alarm outputs (4); binary input (1); USB device port (1), USB host ports (2); Ethernet (1)
- Display 7” WVGA 800x480 pixel, Color LCD w/touch-screen
- Power Universal Power Supply: 100–240 ± 10% VAC, 50 W, 50/60 Hz
- Dimensions H: 54.9 cm (21.6 in); W: 45.0 cm (17.7 in); D: 26.5 cm (10.4 in)
- Weight 15.8 kg (34.9 lb)
- Safety Certifications ETL, CE
- Enclosure Rating IP-45
Consumables
- UV Lamp 6 months
- Acid Reagent As needed, typically 6 months
- Oxidizer Reagent As needed, typically 3-month stability
- Inorganic Carbon Remover (ICR): Improves TOC accuracy for samples with high inorganic carbon levels
- Turbo Mode: Provides four-second analysis time
- DataGuard Software: Full compliance support for 21 CFR Part 11 and Annex 11