H2S/TOS TS MEDORSulfur Compounds Analyzer

SHARE

The H2S/TOS TS MEDOR® is an industrial gas chromatograph for the analysis and monitoring of sulfur compounds in gas: H2S, Mercaptans, Sulfides and total sulfur.

Most popular related searches

According to the level of concentration several versions exist:

  • Standard with the loop 10μL and bypass for the range 0-15ppm of H2S and above 50ppm TS
  • With the loop 50μL for the range 0-3ppm of H2S and > 10ppm TS
  • Automatic sampling using a loop
  • Loop injection by automatic valve on the column
  • Isothermal gas chromatograph
  • Detection of all compounds eluting from the column performed by MEDOR’s wet cell sulfur specific detector
  • Signal provided by electrochemical reaction between the wet cell electrolyte and the sulfur compounds
  • Automatic calibration/validation of the data, with embeded permeation tube (optional)
  • Continuous monitoring with automatic online sampling
  • Analytical performances:
    • Specific and very sensitive to sulfur compounds
    • Results validation by automatic standard injection
    • Long term stability using detector installed in reservoir
  • Extremely low maintenance
    • Very long life time detector, up to 10 years.
    • No need to change the electrolyte for at least 10 years
    • Low gas consumption, can be reduced in option
    • More than 10 years data storage
    • No cylinders required thanks to internal calibration tube and gas generators
  • Automatic control with process device
  • Intelligents system with tunable and interactive alarm levels
  • Powerful VISTACHROM Chromatotec® software:
    • Remote monitoring & injection control
    • Full traceability with on board archiving of results and chromatograms
    • QC Set up and control of threshold alarms
    • Data export by MODBUS / 4-20 mA / 0-10 V
    • Time stamp results

Compounds Analysed:

  • H2S speciation and TOS. TS made by sum

Detection Limit: in ppm

  • H2S : 1.5 ppm (standard version): loop= 10μL
  • H2S : 0.25 ppm: loop= 50μL

Detection Range: H2S / TOS / TS

  • Programmable for H2S: 0/ 05 ppm or 0/ 30 ppm
  • Programmable for TS: 0/ 10 ppm or 0/ 100 ppm or 0/200 ppm
  • Other range upon request (% using specific valve with internal sampling loop)

Relative Standard Deviation:

  • RSD < 3% on concentration over 48H.
  • RSD < 0.6% on retention time over 48H.

Cycle Time:

  • H2S / TS result in 120 seconds for standard version

Storage / Transfer of Results:

  • Hardware storage
  • MODBUS communication protocol (optional)
  • 4-20mA (optional)

Gas supply:

  • Carrier: Dry air or N2 (3 bars): 5 ml/min.
  • CALIB: in continuous 50 ml/min. (option )
  • CALIB during validation ~ 250ml/min (option)

Power supply:

  • Main (230V / 115V 50/60Hz)
  • 24V battery (optional)

Electrical consumption:

  • 150 VA

Dimensions and weight:

  • Rack: 4582 mm 19"
  • Height: 222 mm (5U)
  • Width: 482mm
  • Depth: 600 mm
  • Net weight: 20 Kgs