TECHNOLOGIE MEDICALE
  1. Companies
  2. TECHNOLOGIE MEDICALE
  3. Products
  4. Technologie Medicale - Model REGSON TM2 ...

Technologie MedicaleModel REGSON TM2 -Pressure Regulators

SHARE

The REGSON TM2 pressure regulator, connected to an Oxygen or medical AIR cylinder, is used to regulate the gas pressure by supplying the operating outlet with a controlled and stable pressure below its supply pressure. The pressure of the compressed gas cylinder can be read on the pressure gauge of the device. Combined with a DEBSON TM2 flowmeter with pre-adjusted flowrates, the REGSON TM2 pressure regulator enables to adjust and measure the flow of a gas that is to be delivered to patients through the respiratory tract from a gas cylinder. Fitted with a quick-release connector it also enables to supply a ventilator or any other medical device with gas.

Most popular related searches
  • 2 connection systems to the gas cylinder:
    • Bullnose fitting: for large cylinders (> 4 litres in water or 0.8 m3).
    • Yoke pin-index fitting: universal system for small cylinders (< 4 litres in water or 0.8 m3).
  • Supply pressure: 200 bar
  • Outlet pressure: 4.5 bar
  • Piston type pressure regulator for a great reliability.
  • Single stage pressure regulator.
  • Integrated relief valve protected by the body of the pressure regulator thus ensuring its inviolability and its reliability.
  • Pressure gauge protected against shocks thanks to a plastic bumper.
  • Filter at the inlet to prevent the risks linked to high pressure.
  • It can be used in any orientation for a perfect adaptation to the medical needs especially for emergency.
  • Flow control knob equipped with a ratchet mechanism avoiding accidental flowrate change.
  • Presence of a gas flow between two positions ensuring the gas supply for the patient even if the device is accidentally positioned between two settings.
  • Control knob with large wings facilitating the grip and including a twin display of the selected flowrate.
  • Control knob with a wide transparent window to be able to read the selected flowrate as well as the previous and the next ones.
  • Rotating direction: counterclockwise.
  • Supplied in standard with a Stop system preventing the medical staff from switching from 0 to maximum flow and conversely.
  • Unit serial number engraved on the body of each pressure regulatorensuring its identification and traceability.
  • Device to be serviced every 5 years.

Gases: O2 (Also available in medical AIR for AFNOR, DIN and NORDIC standards)

Flowrates:
1.5 l/min (paediatric) – 5 l/min – 15 l/min – 30 l/min

  • With or without quick-release connector
  • With or without Stop system
  • With or without VENTURI TM2 ejector (OXAP)

Bullnose fitting:

  • O2: AFNOR (French standard), BS (British standard), DIN (German standard), US OHMEDA (American standard), UNI (Italian standard), NORDIC (Scandinavian standard). Contact us for other standard.
  • Medical AIR: AFNOR (French standard), DIN (German standard), NORDIC (Scandinavian standard).

Yoke pin-index fitting: universal system