Zhejiang Tailin Bioengineering Co., Ltd.

TailinModel DI1500 -TOC Analyzer

SHARE
The HTY-DI1500 laboratory TOC Analyzer is designed to determinate the TOC levels of purified water, WFI and Ultra-purified water. It uses UV lamp oxidation technology for accurate, dependable operation with minimal maintenance. The analysis is made by the software installed on the computer, it is easy to operate and in compliance with the GMP and 21CFR PART 11 requirements.
Most popular related searches

Dual-wave UV oxidation with conductivity sensor to ensure reproducibility and long calibration period.
Dynamic detection technology combined with professional algorithms ensures good repeatability.
Dural sensor configuration(TC, IC sensors) eliminates the interference of inorganic carbon during TOC determination.
Observation windows for UV lamp and pump.
Less-disassemble design, easy to maintain.
CSV (Computerized System Validation),21 CFR Part 11 compliance, audit trails compliance.
It meets the regulatory requirements for ChP, USP<643>, EP2.2.44, and JP16.
Match for all the AUTOSAMPLERS made by TAILIN.

  • Power supply: (100-240)VAC 50/60Hz
  • Power consumption: 100W
  • Indication error: ±5%
  • Repeatability: RSD≤2%
  • Determination Range:(0-1500.0)μg/L
  • Test time: 4 min
  • Required sample conductivity:(0-5.1)µS/cm@25℃
  • Required sample temperature:(1-95)℃
  • Ambient temperature:(10-40)℃Dimensions: 23cm x 24cm x 34cm

Validation protocols

System suitability test

Online instructions for software installation