HORIBA Europe GmbH
  1. Companies
  2. HORIBA Europe GmbH
  3. Products
  4. HORIBA - Model HP-300 - Two-Wire ...

HORIBAModel HP-300 -Two-Wire Transmitter

SHARE

Improved single-touch calibration for greater ease of use. Calibration history display. New pH electrode lineup.

Most popular related searches
  • Measuring method : Glass electrode method
  • Measuring range : pH: Oto 14 pH : Resolution: 0.01 pH
  • Temperature: 0 to 1008C : Resolution: 0.1 aC
  • Repeatability : pH: ±0.03 or less : Temperature: ±0.38C (for
  • equivalent input)
  • Linearity : pH: ±0.03 or less : Temperature: ±0.38C (for
  • equivalent input)
  • Transmission output : 4 to 20 mA DC : Input/output insulated type
  • Maximum load resistance 600 Q
  • Contact input : One point : Contact format: Non-voltage a contact
  • Contact function: Transmission output is held for closed contact input.
  • Temperature : Oto 1008C
  • compensation range
  • Ambient temperature : -20 to 608C
  • Temperature : Pt 1000 (08C)
  • compensation element : Positive temperature sensitive resistor element: 500
  • Q (258C), 6.8 kQ (258C), 10 kO (258C)
  • Calibration method : Automatic or manual calibration
  • Self-diagnosis function : Calibration error, Electrode diagnosis error,
  • Transmitter malfunction
  • Power supply : 24 V DC (operational voltage range: 21 V to 32 V
  • DC) Power consumption 0.6 W or less
  • Construction : Outdoor installation type: IP65 protection level Mounting method: 50A pole or wall mounted Case: Aluminum alloy Mount fitting/hood: SUS304
  • Weight : Approx. 4 kg
  • Equipment protected : CE marking, FCC rules