NOVONIX
  1. Companies
  2. NOVONIX
  3. Products
  4. NOVONIX - Model CMA-HDX-99-56 - UHPC 2A ...

NOVONIXModel CMA-HDX-99-56 -UHPC 2A Channel Module

SHARE

NOVONIX's UHPC 2A Channel Module offers precise measurement and control of current and voltage with extremely low noise levels, making it a reliable device for generating high-quality data. It provides a range of current from 200µA to 2A and voltage up to 5V, with resolution and accuracy designed for meticulous scientific experiments. The device incorporates integrated RTD temperature sensing for accurate thermal data and supports multiple channel operations with its 8 test channels. Designed to fit standard lab environments, it operates between 18°C and 28°C with a power draw of 320 VA, accepting input voltages from 120VAC-230VAC. For optimal performance, especially under heating conditions, it is recommended to pair with NOVONIX Thermal Chambers.

Most popular related searches

NOVONIX’s industry leading Ultra-High Precision Coulometry (UHPC) Channel Modules accurately measure and control current and voltage, with high precision and low noise, providing better quality data that allow for greater confidence.

 

Current
Range: 200µA, 2mA, 20mA, 200mA, 2A
Resolution: <20ppm FSR
Accuracy: 0.01% FSR
Noise: 0.002% FSR
Temperature Coefficient: 0.002% FSR over 23 ±5°C

Voltage
Range: 250mV, 5V
Resolution: 1µV
Accuracy: 100µV (<250mV), 200µV (>250mV)
Noise: 100µV
Temperature Coefficient: 0.002% FSR over 23 ±5°C

Current
Resolution: <20ppm
Accuracy: 0.005% FSR
Noise: 0.002% FSR
Temperature Coefficient: 0.002% FSR over 23 ±5°C

Voltage
Resolution: <50nV on 0V-0.25V, 1µV on 0.25V-5V
Accuracy: 100µV (<250mV), 200µV (>250mV)
Noise: 100µV
Temperature Coefficient: 0.002% FSR over 23 ±5°C

Resolution: 10ms
Measurement Frequency: 6Hz
Accuracy: 2ppm