Mettler - Toledo Int. Inc
  1. Companies
  2. Mettler - Toledo Int. Inc
  3. Products
  4. UniCond - 2-Electrode Conductivity ...

UniCond2-Electrode Conductivity Sensors

SHARE

UniCond sensors contain electronics embedded in each sensor. This innovative technology allows a very wide measurement range: a single UniCond 2-E sensor covers a range from ultrapure to brackish water. METTLER TOLEDO’s proprietary Intelligent Sensor Management (ISM) concept adds further benefits. ISM converts the conductivity measurement into a highly stable digital signal which is output to the transmitter. In addition, ISM`s Plug and Measure feature means sensor installation is fast and easy.

Most popular related searches

The UniCond 2-E series is suitable for measuring conductivity in pure and ultrapure waters, pharmaceutical waters, make up water, and wastewater.

Wider rangeability with titanium and Monel sensors -  ultrapure water to 50,000 µS/cm

Stainless steel sanitary Tri-Clamp sensors - 0.02 to 2,000 µS/cm total range with +/- 1% system accuracy easily meets pharmaceutical water requirements

Specifications - UniCond 2-electrode

Short description: Digital sensor measures from ultrapure to brackish water.
Accuracy (cell constant): 0.01 cm-1 sensor: ± 1%0.1 cm-1 sensors: ± 1% for 0.02-5,000 µS/cm; ±3% for 5,000 - 50,000 µS/cm
Temperature sensor: Pt 1000 RTD, IEC 60751, Class A, with NIST-traceable calibration
Temperature accuracy: ± 0.1 °C at 25° C
Connector: IP65, mates with 58 080 27X series cable

Features and Benefits

  • Extended conductivity measurement range – one sensor covers UPW to brackish water
  • 33% improved accuracy over analog sensors
  • Digital signal ensures high process reliability in harsh environments
  • No signal degradation over long cable lengths
  • Highly repeatable accuracy due to patent pending calibration system
  • Can be calibrated in-line as required by USP
  • Installation convenience with Plug and Measure
  • Identification, calibration and maintenance data stored in the sensor for easy traceability
  • Measuring circuit built into sensor body eliminates cable resistance and capacitance effects
  • NIST traceable resistances to cover ranges of measurement—conductivity and temperature