Sartorius AG
  1. Companies
  2. Sartorius AG
  3. Products
  4. Sartorius - Model CCL1007 - Vacuum ...

SartoriusModel CCL1007 -Vacuum Comparator

SHARE

Vacuum mass comparator for 1 kg mass prototypes of national metrological institutes for research and for highly accurate mass calibration. The electronic vacuum mass comparator features a vacuum chamber and load alternator for 1 kg mass prototypes and is capable of determining differences in mass to an accuracy of 0.1 μg for weights of 1 kg - even under high-vacuum conditions.

Most popular related searches
  • The comparator is protected from environmental effects by a pressure-stable, non-magnetic aluminum chamber.
  • The unique three-point holder is capable of handling both cylindrical weights and silicon spheres efficiently.
  • The weigh cell is located in the upper section of the vacuum chamber. The weighing pan for the substitution weights is accessed via the upper loading hatch.
  • The motors that drive the load alternator and the weight changing device are located outside the vacuum chamber. This reduces interference caused by temperature fluctuations.
  • All components inside the vacuum chamber are completely resistant to high vacuum and, therefore, will not outgas under vacuum. This prevents any contamination of the weights.
  • The more than 20 vacuum flanges enable additional sensors to be introduced into the vacuum chamber, which has sufficient space to accommodate additional measuring instruments.
  • Inclusive Printer and PC

Sartorius has mastered the core disciplines of weighing like no other company. Sartorius sets new standards in mass metrology. In cooperation with the Bureau International des Poids et Mesures and the Institute for Process Measurement and Sensor Technology of the Technical University of Ilmenau, Sartorius has developed a mass comparator — the CCL1007 — that is capable of determining differences in mass to an accuracy of 0.1 μg for weights of 1 kg — even under high-vacuum conditions.

Product Properties

  • Maximum capacity: 1,031 kg
  • Readability: 0,1 µg
  • Repeatability (typical), s: 0,1 µg

Technical Attributes

  • Application Area: 1 mg - 1 kg
  • Repeatability: 0,2 µg
  • Repeatability in vacuum: 0,1 µg
  • Electronic weighing range & taring range: 2 g
  • Linearity: 1 µg
  • Off-center loading error: 0 µg
  • Range sensitivity: 0,2 µg / 500 mg
  • Load alternator: 8
  • Typical stabilisation time: 80 s
  • Cycle time ABBA: 480 s
  • Pressure range: 10E-6 – 1.000 mbar
  • Centermatic: Yes
  • Control unit: Yes
  • PC Software: Yes
  • Weighing pan diameter: 50 mm
  • Sample size (WxD): 34 - 95 x 110 mm
  • Diameter range for a silicon sphere: 45 - 100 mm
  • Helium leakage rate: < 10E-7 (mbar*l)/s