VIG Industries - Model 200 FID Dual Channel - Oven Heated Total Hydrocarbon Analyzer
From VIG Industries - Flame Ionization Detector (FID) Hydrocarbon Analyzers
The VIG Industries, Inc. Model-200 is a microprocessor based, oven heated methane / non methane / total hydrocarbon gas analyzer designed for high accuracy, sensitivity and stability. The Model-200 uses two independent flame ionization detectors (FID), one to measure total hydrocarbons and the second coupled with a GC (gas chromatography) column for the separation of the methane and non methane components. A sample is fed to the analyzer via an internal heated pump to the first FID for a continuous real time total hydrocarbon reading
-
Most popular related searches
- total hydrocarbons analyzer
- flame ionization detector
- hydrocarbons analyzer
- volatile organic compounds
- total hydrocarbons gas analyzer
- non-methane total hydrocarbons
- heated total hydrocarbons analyzer
- heated sampling pump
- flame ionization detector hydrocarbons analyzer
- non-methane total hydrocarbons analyzer
Applications
- Compliance Monitoring - US EPA Method 18 and Method 25A
- Process Monitoring - Continuous monitoring and alarm or control of:
- Process gas streams utilizing organic solvents
- Crude oil
- Other chemicals containing hydrocarbons
- Efficiency Monitoring - Monitoring effluent of volatile organic compound (VOC) reduction equipment for:
- Environmental compliance
- Efficiency control of incinerators (Thermal or catalytic)
- Scrubbers
- Carbon absorbers
- Monitoring of catalytic converters
- Combustion and diesel engine efficiency
- Other abatement equipment
- Safety Monitoring - Lower explosive limit (LEL) monitoring and/or control of:
- Ovens/dryers
- Fugitive emissions monitoring
- Personnel work area monitoring
- Leak detection of process equipment or solvent storage areas
- Stack Monitoring
Features
- Easy to use software
- Gas chromatography GC technology used for separation
- Automatic fuel shut-off system
- Automatic start-up/ignition
- Automatic flame-out indicator
- Adjustable alarm and oven settings
- Precision 1% of full scale
- Fast response
- Methane and Non-methane - minimum 2.5 minute cycle
- Total Hydrocarbons - Within 5 seconds to 90% of final reading (continuous real time reading)
- Two stage sample filter with exchangeable sintered stainless steel elements
- Teflon isolated detectors (FID)
- 19' rack or bench mount
- Heated Sample pump heads
- Outputs 0-1, 0-5, or 0-10Volts
- Automatic ignition
Options
- Zero and calibration input ports with software
- 4-20mA output - no extra charge
- Range recognition relays
- RS-232 interface
- Internal combustion air supply
- Automatic sample shut-off system
Related Available Equipment
- External zero air generator (reduces bottles)
- Hydrogen Generator (reduces bottles)
- Heated sample lines and controllers
- Strip chart recorders and data loggers
- NEMA rated enclosures
Standard Specifications
- Analysis Method: 2 Oven Heated, Flame Ionization Detector (FID)
- Separation Method: GC Column
- Precision: +/- 1% of Full Scale
- Response Time: Total Hydrocarbons - Within 5 seconds to 90% of final reading (continuous real time reading). Methane - Approximately 40 seconds, updated every 3 minutes. Non-methane - Approximately 70 seconds, updated every 3 minutes.
- Zero Drift: +/- 1% of Full Scale Per 24 Hours
- Span Drift: +/- 1% of Full Scale Per 24 Hours
- Linearity: Within 1% of full scale
- Oxygen Synergism: Within 1% of full scale within selected range
- Ranges: (4 Ranges per amplifier, two amplifiers per analyzer, one amplifier for total hydrocarbons, one amplifier for methane and non-methane) Standard Ranges 0-100, 1000, 10000, 100000ppm (LDL 0.1ppm). 0-10, 100, 1000, 10000ppm (LDL 0.01ppm).
- Display: Graphic, backlit, 240W x 64H pixels, high contrast, wide viewing angle
- Outputs: 0-1, 5, or 10 Volts & optional 4-20mA
- Alarms: Flame Out & Concentration, Latching
- Safety: Flame Out Indicator with Fuel Shut-Off System
- Oven Temperature: Operating Temperature 275°F, Adjustable 200 to 300° F
- Sample Pump: Heated 4 Liters/Minute standard or Heated 14 Liters/Minute optional
- Ignition: Automatic (can be set to manual by operator)
- Glow Plug : Main and spare glow plug, selectable by switch on back panel
- Dimensions: 19' front panel, 16.75' wide chassis, 18' deep chassis, 21' deep with fittings and handles, 9' high
- Weight: 55 to 65 pounds depending on options
Operating Requirements
- Fuel: Hydrogen (UHP Grade) 18psi incoming pressure
- Combustion Air: Zero Grade Air 18PSI
- Operation Air: Compressed Air (Instrument Air - Dry / Oil-Less) 60PSI
- Zero Gas: Zero Air (UHP Grade) 10PSI
- Calibration Gas: User Selected 10PSI
- Carrier Gas: Nitrogen (UHP Grade) 30PSI
- Power: 115Vac, 60Hz or optional 220Vac, 50Hz
- Ambient Temperature: 32 to 110º F