Xonteck Inc
- Home
- Companies
- Xonteck Inc
- Products
- Xonteck - Model 910 - Canister Air ...
Xonteck - Model 910 -Canister Air Sampler
FromXonteck Inc
The Xonteck Model 910 Canister Sampler is a new features manual sampler that is designed to collect volatile organic compounds in ambient air. The method is based on collection of whole air samples into `Summa` electropolished canisters as outlined in USEPA TO-14A/TO-15 Methods.
Most popular related searches
mass flow control
air sampling canister
air sampling
canister sampler
flow control
digital panel meter
volatile organic compounds
whole air sampling
pressure gauge
mass flow
- Low cost – manual sampler
- Meets all PAMs and TO-14A/TO-15 requirements
- Convenient display layout for manual data recording
- Mass flow controller maintains constant and adjustable flow rate
- Digital display of flow rate
- Stainless steel, Teflon® and Viton® wetted surfaces, certified clean*
- Automatic system purge prior to sampling
- Easy installation, operation and maintenance
- Sample Flow Control: Porter Model 201, Mass Flow Controller, typically 0-20,0-50,0–100sccm or customer specified, accuracy and linearity to ±1% of full scale.
- Sample Flow Adjustment: precision potentiometer with turning number indicator, maximum turns 10, resistance range 0Ω - 5kΩ, tolerance to ± 5 %.
- Display: Digital Panel Meters, 0.56” LED display.
- Sample Hours Display: time totalizer, 8 digit LCD display with 999999.99 hour capacity, resolution 0.01 hour.
- Timer: 24 hour, 7 days electronic timer module with switching status indication, LCD display, 0.35”.
- Internal Sample Pump: AD B161 diaphragm pump, SS single head, max pressure 29.2 psig, max vacuum 23.0”Hg, max flow 7.6 lpm.
- Pressure Gauges:
- Canister Pressure: U.S. Gauge, dial size 2-1/2”, range 30”Hg-0-30 psig, minor graduation 2”Hg(30”Hg-0), 1psig(0-30psig), accuracy ±3-2-3% of span (Grade B).
- Pump pressure: U.S. Gauge, dial size 2”, range 0-30 psig, minor graduation 1 psig, accuracy ±3-2-3% of span (Grade B).
- Pump Pressure Adjustment: Swaglok SS Medium-Flow Angle-Pattern Metering Valve, 1/8” end connection.
- Connections:
- Inlet: 1/4” tube fitting
- Outlet: 1/8” tube fitting
- Bypass: 1/8” tubing fitting
- Power: 115V/60Hz; 220V/50Hz(optional).
- Dimensions: 7”(H)×19”(W) ×13”(D)
- Weight:19lbs