Cerulean
  1. Companies
  2. Cerulean
  3. Products
  4. Cerulean - Model FPS-1 - Air Sampler ...

CeruleanModel FPS-1 -Air Sampler Calibration Check Device

SHARE

Cerulean Air Sampler Calibration Check Device is designed to check the calibration of Cerulean AS600 and AS600XR Air Samplers. At the heart of Cerulean Air Samplers is a flowmeter that uses a differential pressure orifice plate. As air flows through the orifice plate, it creates a pressure differential between the two faces of the  plate.  The difference in  pressure is dependant upon the flow rate and  thus accurate measurement of this pressure differential determines the flow rate. The pressure differential is detected by two pressure transducers located in the instrument box of the Cerulean Air Sampler. 

Most popular related searches

They are  attached to the orifice  plate by two translucent pneumatic tubes. The Cerulean Air Sampler Calibration Check Device operates by applying the same  pressure to one of the pressure transducers (the other being left open to atmosphere) and the manometer in the calibration check device. Readings are taken from the Air Sampler and the Calibration Check Device. The gauge of the manometer is converted to a flow rate so that it can be easily compared to the Air Sampler.

  • Easy to use calibration check device
  • No calibration required
  • Allows fast verification of air flows on Sampler Cerulean AS600 Air
  • Allows fast verification of air flows on Sampler Cerulean AS600XR Air
  • Cerulean offer a recalibration service for the AS600 and
  • AS600XR Air Samplers with certificate if required
  • Used for checking the calibration of Cerulean AS600 and
  • AS600XR Air Sampler units.