The Transparency Analyzer is connected directly via RS485 communication interface, providing simple, reliable, cost-saving process data with remote monitoring, calibration, configuration and diagnostics capabilities. Transparency Meter in a robust IP68 proof enclosure, 1500 N tensile strength Kevlar reinforced cable, up to 1.2 km digital data transmission, the transmitter is ideally used in water/wastewater industry.
Most popular related searches
water monitoring method
surface water monitoring
surface cleaning
remote monitoring
wastewater industry
surface water
water monitoring
water enclosure
source sampling
water industry
- Robust IP68 Water Submersible Enclosure, Solution Without Cabinet
- On-line Realtime Measurement
- Ultra Low Power Consumption, Ideal for Outdoor Applications
- High Reliability, Drift Free
- Surge Protection for Power and RS485 Communication
- 1.2 Km RS485 Digital Communication, Minimize Cabling and Engineering Cost
- Software Configuration Calibration and Data Monitoring
- Standard Modbus RTU Protocol, Direct Connected with PLC, HMI, Eliminate I/O Module Cost
- Sapphire Glass Windows to Prevent Scratch
- Innovative Nano Coating to Remain Window Clean
- Optional Auto Cleaning Wiper, Almost No Maintenance
- Onboard memory allowing users to easily calibrate and configure sensor at lab and distribute to various fields sites
- Measurement range: 0~2 m
- Accuracy: ±5% FS
- Resolution: 1 mm
- Repeatability: ±3% FS
- Light source: LED 880 nm
- Beam angle: 180°
- Process flow rate: 0.1~10 m/s
- Operation pressure: Max.5 Kgf/cm²
- Operating temperature: 0 ~ 45 °C
- Response time: 1 sec
- Protection: IP68
- Housing: SS316L
- Dimension: Φ 39Χ344 mm
- Weight:
- analyzer: approx. 0.7 Kg
- Cable: 80g/m
- Output Signal: RS485 (Modbus RTU protocol), 19,200 bps, 8 data bits, no parity, 1 stop bit; 4~20 mA (optional)
- Data Resolution: 16 bits (0.001% FS)
- Surge Protection: 4000 V DC
- Power: 3.6~12 VDC ±10%, 45 mA
- Protection: polarity, overload, short circuit
- Safety: CE, FCC
- Surface water
- The analyzer consists of a light source, a sample cell, and a light detector (photo detector). Incident light is absorbed by the sample, and the light is measured by the detector. The value depends on the amount of absorbance in the sample. The transmitter uses a long life LED and the 180° light method. An automated mechanical wiper is to remain surface clean and remove air bubbles of the optical window in order to maximize the accuracy and minimize the maintenance requirement.