The Epsilon EClipse is the latest electrochemical analyzer manufactured by BASi. It is a potentiostat/galvanostat with a second working electrode for bipotentiostat measurements. Software with the most popular electrochemistry techniques is included with the instrument. 

Most popular related searches

The Epsilon EClipseTM is the latest electrochemical analyzer manufactured by BASi. It is a potentiostat/galvanostat with a second working electrode for bipotentiostat measurements. Software with the most popular electrochemistry techniques is included with the instrument.

This System Includes:

  • Basic Software - techniques include
    • Cyclic Voltammetry/Linear Sweep Voltammetry
    • Chronoamperometry/Chronocoulometry
    • Controlled Potential Electrolysis
    • DC Potential Amperometry
    • Open Circuit Potential vs. Time
    • Chronopotentiometry
  • Basic Plus Software - techniques include
    • Sampled Current Polarography
    • Normal Pulse Voltammetry/Polarography
    • Differential Pulse Voltammetry/Polarography
    • Square Wave Voltammetry
    • Stripping Voltammetry Techniques
    • Double Potential Step Chronopotentiometry
  • Methods Software - techniques include
    • Prepare sequences with an unlimited number of techniques
    • Insert trigger and delays in sequence
    • Repeat sequence up to 9999 times
  • Bi-Potentiostat - techniques include
    • Multi-Channel Amperometry
    • Multi-Channel Cyclic Voltammetry 
    • Multi-Channel Chronoamperometry

Potentiostat

  • Channel 1 applied potential: Dynamic DAC: 16 bit, ±3.275 V @ 0.1 mV resolution or ±10 V @ 0.33 mV resolution
  • Offset DAC: 8 bit, ±2.55 V @ 10 mV resolution
  • Channel 2 applied potential: Dynamic DAC: 8 bit, ±2.55 V @ 10 mV resolution
  • Compliance voltage: ±12 V
  • Maximum current: 100 mA
  • Bandwidth: > 100 kHz
  • Bias current: < 1pA
  • Electrometer input impedance: > 1012W
  • (values for parameters other than applied potential are for both channels)

Current-to-voltage converter

  • Full scale sensitivity: 1 nA*, 10 nA*, 100 nA, 1 µA, 10 µA, 100 µA, 1 mA, 10 mA,100 mA, (*secondary gain used)
  • Secondary gain: x1, x10, and x100
  • ADC resolution: 16 bit
  • Sampling rate: 50 kHz (20 µs/conversion)
  • Data length: up to 64k points - fast

Galvanostat

  • Applied current: 50 pA - 50 mA
  • Range: ±10 V ± 1 V
  • Measured potential resolution: 0.02 mV
  • Max leakage current: 30 pA

I/O

  • Eout (applied potential)
  • Iout (current to voltage output)
  • Ein (input connection for external applied wave form)
  • Cell Stand port (control of C-3 Cell Stand and CGME)
  • Ordering Port (control of RDE-2)
  • PC (USB) port
  • Trigger out (TTL)

Minimum PC requirements

  • Win 7 or higher
  • Pentium III
  • 512 MB RAM
  • 50 MB hard drive space available
  • USB port

Power requirements

  • 120 or 240 VAC, 50/60 Hz, 60 VA

Dimensions & weight

  • 15.75" (40 cm) x 5.25" (13 cm) x 12.75" (32.5 cm)
  • 17.5 lb (7.4 kg) (1 channel)