ENVEA
  1. Companies
  2. ENVEA
  3. Products
  4. ENVEA - Model AS32M - CAPS Nitrogen ...

ENVEAModel AS32M -CAPS Nitrogen Dioxide Analyzer NO2

FromENVEA
SHARE

Gas monitor for fast, precise, direct and continuous measurement of NO2 concentrations in the range 0-1000 ppb. Visible (450 nm) absorption measurement using patented Cavity Attenuated Phase Shift (CAPS) technology. QAL1 certified by the TÜV, US-EPA approved. Fast and direct measurement of the sample – no chemical conversion required, no Toxic Gas Emissions”.

Most popular related searches
  • Essentially interference-free, insensitive to high fluctuations of nitric oxide, aerosols, humidity and other trace atmospheric species
  • Ideal for a wide variety of monitoring application where accuracy, precision and fast response are required
  • Extremely compact, easy to use with minimal maintenance
  • Built-in USB port, serial interface (RS 232 / RS 422) and TCP-IP connection
  • Real-time calibration graph, animated synoptic, control and maintenance data screens can be displayed while the instrument is operating
  • Cost-saving analyzer with ultra low power consumption, 24 V supply (option) and able to withstand temperatures up to 45-50 °C without air conditioning
  • Includes embedded Communication Protocol for XR® Management with automatic recognition & configuration
  • SmartStatusLightTM power button on the front panel indicating status of operation (ON/OFF, Alarm, Maintenance required…)
  • Continuous indoor and outdoor air quality monitoring
  • Stationary and mobile AQMS laboratories
  • Industrial fence-line monitoring
  • Background urban or rural, Traffic, Kerbside measurement campaigns and monitoring studies
  • Roadside air pollution and street canyon effect
  • Tunnel Monitoring, Traffic stations…
  • Measured parameter: NO₂
  • Technology: CAPS (Cavity Attenuated Phase Shift Spectroscopy)
  • US-EPA Approval / QAL 1 Certification: YES
  • Measurement ranges: 0-1 ppm
  • Measurement units: ppb, ppm, µg/m3 (user-selectable)
  • Lower Detectable Limit (2σ): 0.1 ppb
  • Noise (σ): 0.05 ppb
  • Zero drift: 0.75 ppb/24h
  • Span drift: 1.5 ppb/12h
  • Linearity: < 4%
  • Sample flow rate: 1L/min
  • Sampling pump: Internal
  • Temperature range: +10°C to 35°C
  • Data storage: 1 year
  • Included I/O: 1 LAN / 1 USB / WIFI / 1 RS4i board
  • Weight (Kg): 12.5
  • Dimensions mm (LxWxH): 483 x 606 x 133
  • Chassis: 19″ Rack, 3U
  • Power supply: ‘100~250Vac, 50/60Hz + ground

Main Options:

  • TFT color touch screen (800 x 480 , 7″)

Compliance With:

  • 2008/50/EC, EN14211:2012, EN15267-1:2009
  • EN15267-2:2009, 40 CFR PART 53 & 40 CFR PART 58
  • QAL 1 certified by the TÜV to N° 0000038507
  • U.S. EPA approved as compliant with EQNA-1013-210