Dado Lab SRL
  1. Companies
  2. Dado Lab SRL
  3. Products
  4. Dado - Model HP1 - Compact and ...

DadoModel HP1 -Compact and Lightweight, Heated Gas Probe

SHARE

Compact and lightweight, the Dado Lab HP1 heated probe is the versatile solution for gas sampling, fully compliant with EN 12619, EN 14792, EN 14789, and CEN/TS 17405 standards. Designed to adapt to any requirement, it is available in single or dual outlet configurations, with either self-regulation at 180?°C or adjustable temperature control with a freely settable set point. You can choose from three protective filter options, and for the most demanding environments, pair it with a heated sampling line to prevent acid condensation and ensure sample integrity. 

Most popular related searches

An LED indicator signals when the set point temperature has been reached. 

Connections are made of stainless steel with Swagelok-type fittings, available in 6 mm or, upon request, 8 mm diameters, specifically selected to withstand prolonged thermal stress. 

The filter element is housed in an easily removable holder with a very low dead volume, reducing the time needed for the sample to reach the analyzer or sampling line. 

The probe also features a second inlet for the dilution line, positioned upstream of the filter chamber to facilitate gas mixing. This inlet can be used for zeroing or calibrating analyzers, or for creating dilution lines via the QB1-D sampler. 

To avoid cold spots, the HP1 is equipped with the DadoLab Locking Ring, securing the heated line and preventing temperature drops at the connections. 

Unique in its class, the HP1 probe has been developed to connect to the gas line of the HP5 isokinetic probe, enabling simultaneous operation during isokinetic particulate sampling. 

General

  • Sampling probe: AISI 316 stainless steel, length 300?mm or 700?mm, diameter 8/6?mm (optional titanium tube).
  • Filtering element: Microfibers with PTFE binders protection, Glass fiber or Sintered steel
  • Heating temperature: Self limited to 180°C or thermoregulated
  • Hysteresis: +/- 15°
  • Heated line connection: Dadolab locking ring with Swagelok-type fittings in 6?mm or 8?mm sizes
  • Operative conditions: -20 ÷ 40°C – 95% rH
  • Stock Conditions: -10 ÷ 50°C – 95% rH
  • Power Supply: 230/110 Vac ±10% 50/60Hz
  • Materials: Stainless steel
  • Weight: 2,3 kgs
  • Dimensions: 1570 (+300) x 94 x 134mm