- Home
- Companies
- Hiden Analytical Ltd.
- Products
- Hiden - Model HPR-20 EGA - Compact ...
Hiden - Model HPR-20 EGA -Compact Benchtop Gas Analysis System
A compact bench-top gas analysis system for evolved gas analysis in thermogravimetric mass spectrometry, TGA-MS. The Hiden HPR-20 EGA gas analysis system is configured for continuous analysis of evolved gases and vapours from thermogravimetric analysers, TGA. The HPR-20 EGA system includes EGAsoft PC software, an application specific software program designed for TGA-MS, including data export in formats to suit direct import of MS data into the TGA equipment software.
The Hiden HPR-20 EGA is a bench top mass spectrometer with QIC (quartz inert capillary) that samples from 100 mbar to 2 bar gauge and is included as standard.
The HPR-20 EGA system includes an external dry scroll pump to provide for high performance combined sample bypass and UHV foreline pumping, providing fast response and effective pumping for light gases.
Effective pumping for light gases enables the use of helium as the carrier gas for TGA-MS studies without compromise to sample flow and response. The benefit of using helium as a carrier gas of choice for TGA-MS studies is the reduced mass spectral interference with helium, compared to nitrogen or argon for example.
The HPR-20 EGA system has mass range 200 amu (300 amu option) and with a detection capability from 100% to less than 100 part-per-billion.
An extensive range of MS inlets for optimised coupling the HPR-20 EGA mass spectrometer to TGA systems is offered to suit an existing TGA, or OEM partner TGA equipment.
The TGA interface includes re-entrant furnace sampling, providing close coupling to the TGA furnace region for optimized evolved gas/vapour analysis.
Custom designed interfaces are available for special requirements with alternative systems being offered for applications requiring direct sampling from advanced thermogravimetric analysers operating at higher pressures to 30 bar.
- Enhanced pumping for light gases
- Heated inlet for non-tailing response to desorbed gases and vapours
- Control of ionisation energy parameters for simplification of cracking patterns
- Custom, low dead volume interface to specific manufacturers’ TGA systems
- Optimised data acquisition using EGAsoft
- Mass range option 300 or 510 amu
Configured for TGA-MS continuous evolved gas analysis, the HPR-20 EGA exemplifies the Hiden Analytical approach to gas analyser manufacturing and supply. See the full instrument specifications below:
- Mass range: 200 or 300 amu
- Minimum detectable concentration: 100 PPB
- Response time to gas concentration change: 300 milliseconds
- High capacity vacuum pumping: Yes
- Sample inlet: Heated flexible silica capillary
- Standard capillary length: 2 meters
- TA-MS adaptors/interfaces to suit most TGA systems: Yes
- Carrier gas: Helium or Argon
- Sample flow required (Standard): 16 atm cc/min
- Sample flow required (Optional): Options for < 1 atm cc/min to > 20 atm cc/min
- Fast measurement speed: Up to 650 measurements per second
- Dedicated evolved gas analysis software: Yes
- Designed for TGA-MS: Yes